EtherSAM Tabs
344
IQS-8500 Series and IQS-8120NGE/8130NGE
Overview (Results)
Alarms
Link Down
: Indicates that the Ethernet connection is down due to
a local or a remote fault condition.
LOS
(Loss Of Signal): A
LOS
indicates that there is no optical input
signal.
Frequency Alarm
: The frequency alarm indicates that the received
signal rate meets the standard rate specifications (green) or not
(red). For electrical port, refer to
optical ports, refer to
on page 140. The frequency alarm
will not be considered in the global verdict.
LOPPS-L
and
LOPPS-R
(Loss Of Pulse Per Second - Local/Remote)
is declared when either no pulse is received or when no pulse is
received within 1 second ± 6.6
μ
s after the previous pulse.
LOPPS-R is only monitored once the DTS connection is
established.
Service Configuration Test Overview
Displays an overview of the Service Configuration Test results. The
following fields are presented in the same order as they appear in the
table:
Services No
indicates the service number.
Dir
.: Direction is displayed for dual test set test case providing
bidirectional data.
L -> R
indicates results from local to remote direction.
R -> L
indicates results from remote to local direction.
Frame Loss (%)
indicates the percentage of frames that are lost. Note
that the reported value is the maximum percentage of Frame Loss
from all the burst sequences and ramp steps excluding the
CIR+EIR
,
EBS
, and
Traffic Policing
steps.
Max Jitter (ms)
indicates the maximum measured delay variation.