Expert Mode Tabs
438
IQS-8500 Series and IQS-8120NGE/8130NGE
Port (BERT)
Frequency
Enter the minimum and maximum frequency offsets either in ppm or bps
according to the unit selection.
Frequency (bps)
: Indicates the frequency of the input signal.
Offset Max
: Enter the frequency maximum value allowed before
declaring a Fail condition. Choices are from
0
to
120
ppm. The default
setting is
100
ppm.
Offset Min
: Enter the frequency minimum value allowed before
declaring a Fail condition. Choices are from
-120
ppm to
0
ppm. The
default setting is
-100
ppm.
Status
: Indicates the real-time pass/fail condition.
Grey LED
: Indicates that Expert is disabled or the test did not run yet.
Green LED
: Indicates that the threshold is respected (Pass).
Red LED
: Indicates that the threshold is not respected (Fail).
Unit
: Select the unit. Choices are:
bps
and
ppm
. The default setting is
ppm
.