SmartR™ Features
Copper and G.fast/VDSL2/ADSL2+ Multi-play Test Set
159
Pair Detective
To set threshold values:
1.
Press the up/down arrow keys to highlight the desired parameter.
2.
Press
on a value to open the alphanumeric editor screen and use
the navigation keys to scroll through.
3.
Press
to select each value and use the function keys to complete
your selection.
4.
Press
to confirm and close your selections.
Multimeter
The
Multimeter
page displays the raw values used by
Pair Detective
to identify a fault on the summary
screen. Read-only results and the
P/F
(pass/fail) status
are shown for the following tests:
DC Voltage
displays the volts for each DC
Pair
.
DC Current
displays the
DC
current results for
T-R/A-B wire
Pair
.
Capacitance/Opens
displays
Capacitance
and
Length
for each wire
Pair
as well as
Capacitive
Balance
among
All Pairs
in
%
.
Isolation
displays
Resistance
and a
Soak Counter
for each wire Pair.
Summary of Contents for MAX-635G
Page 1: ...User Guide Copper and G fast VDSL2 ADSL2 Multi play Test Set MAX 635G www EXFO com ...
Page 10: ......
Page 26: ......
Page 162: ......
Page 218: ......
Page 270: ...TDR 260 MAX 635G Continuous Examples of Common Faults Short Fault ...
Page 271: ...TDR Copper and G fast VDSL2 ADSL2 Multi play Test Set 261 Continuous Open ...
Page 272: ...TDR 262 MAX 635G Continuous Bridged Tap ...
Page 273: ...TDR Copper and G fast VDSL2 ADSL2 Multi play Test Set 263 Continuous Splice Joint ...
Page 274: ...TDR 264 MAX 635G Continuous Peak Function ...
Page 275: ...TDR Copper and G fast VDSL2 ADSL2 Multi play Test Set 265 Continuous Dual Trace View ...
Page 276: ...TDR 266 MAX 635G Continuous Load Coil ...
Page 290: ......
Page 336: ......
Page 414: ......
Page 424: ......
Page 432: ......
Page 438: ......
Page 444: ......