Copper Auto Tests
Copper and G.fast/VDSL2/ADSL2+ Multi-play Test Set
317
User Auto Test
Recalculate
button allows you to recalculate data rates after editing
the following
Test Parameters
:
SNR Margin
is the signal-to-noise ratio margin measured on the
line.
Bits/Bin
allows you to select the ADSL bandwidth bit rate.
Technology
allows you to select a frequency band from a list of
possible technologies (subject to the
Software Options
enabled).
Note:
VDSL2-35b requires the VDSL2-35b Data Rate Prediction software option
and CQ Board Revision J or higher.
Reference Point (side)
is either
CO
(default) or
CPE
. Set this value
based on where you connect the tester: for example, if the tester is
connected at the user side and performs a test toward the CO, set
the
Reference Point
to
CPE
.
Information
Annex
displays the mode:
A
or
B
.
Remote Noise Level
ranges from -40 to -110 dBm/Hz (default) and
is configurable in profile *.cqcfg files. The noise level value is
dependent on the
Reference Point
.
To recalculate the DRP:
1.
Press the up/down arrow keys to highlight the desired parameter.
2.
Press the left/right arrow keys to increase/decrease the values.
OR
3.
Press
on a value to open a list box of options or the alphanumeric
editor screen and use the navigation keys to scroll through.
4.
Press
to confirm the value.
5.
Press the up button to select the
Recalculate
button.
Summary of Contents for MAX-635G
Page 1: ...User Guide Copper and G fast VDSL2 ADSL2 Multi play Test Set MAX 635G www EXFO com ...
Page 10: ......
Page 26: ......
Page 162: ......
Page 218: ......
Page 270: ...TDR 260 MAX 635G Continuous Examples of Common Faults Short Fault ...
Page 271: ...TDR Copper and G fast VDSL2 ADSL2 Multi play Test Set 261 Continuous Open ...
Page 272: ...TDR 262 MAX 635G Continuous Bridged Tap ...
Page 273: ...TDR Copper and G fast VDSL2 ADSL2 Multi play Test Set 263 Continuous Splice Joint ...
Page 274: ...TDR 264 MAX 635G Continuous Peak Function ...
Page 275: ...TDR Copper and G fast VDSL2 ADSL2 Multi play Test Set 265 Continuous Dual Trace View ...
Page 276: ...TDR 266 MAX 635G Continuous Load Coil ...
Page 290: ......
Page 336: ......
Page 414: ......
Page 424: ......
Page 432: ......
Page 438: ......
Page 444: ......