Multimeter Tests
MaxTester Copper Test
107
Resistive Balance
To select the parameter values:
1.
Press the up/down arrow keys to highlight the desired parameter.
2.
Press the left/right arrow keys to view and select the options.
OR
3.
Press
on a value to open a list box of options or the alphanumeric
editor screen and use the navigation keys to scroll through.
4.
Press
to confirm the value.
Resistive Balance
The
Resistive Balance
test measures and compares the resistance of each
leg. This test requires you to short Tip and Ring (T-R) to ground at the far
end (T-R-G short) before proceeding with the test.
Snapshot
The resistive balance
Snapshot
shows results for
R-Leg
,
T-Leg
,
Loop
resistance, and the pass/fail status
for
Resistive Balance
in
Ω
and %. Parameters are as
follow:
Current Group
is the current test group name, for
example
Multimeter
.
Current Test
displays the test being measured.
When FED is
Enabled
, the FED icon and numeric
state are displayed beside the
Current Test
.
Here you can switch to another test from the given
list. However, this action will terminate the
ongoing measurement test.