Multimeter Tests
MaxTester Copper Test
121
Stressed Balance
Stressed Balance
The
Stressed Balance
is an option enabled test that
continuously generates high voltage to the cable pair
under test. The high voltage
stresses
the leads in order
to reveal imbalances that may otherwise be hidden
with the normal VF Balance test.
Current Group
is the current test group name, for
example
Multimeter
.
Current Test
displays the test being measured.
Here you can switch to another
Multimeter
test
from the given list. However, this action will
terminate the ongoing measurement test.
Noise Filter
is a fixed value set to
C-message
.
Termination Resistance
is a read-only value of
600
Ω
.
Threshold Settings
button opens a new page that allows you to set the
Stressed Balance Threshold
for the test.
Result
displays the measurement value for
Stressed Balance
and
pass/fail status dependent on the
Good
/
Marginal Balance
threshold
settings.
Test Status
displays the current state of measurement or an error
message.
To select the Stressed Balance parameter values:
1.
Press the up/down arrow keys to highlight the desired parameter.
2.
Press the left/right arrow keys to view and select the options.
OR
3.
Press
on a value to open a list box of options.
4.
Press
to confirm the value.