Frequency Tests
176
MAX-610
VF/AC Balance
Active Result Details
The
Active
Results Details
read-only page displays
the
VF/AC Balance
value and the pass/fail status.
Passive
The
Passive
function computes the passive balance
as the difference between
Power Influence
and
VF
Noise
, displaying and updating results in
real time
,
until you stop the test. The last valid results then
remain displayed. The measurement results of the
active and passive VF Longitudinal Balance tests will
generally be different due to the fact they use different
measurement techniques.
Current Group
is the current test group name, for
example
Frequency Tests
.
Current Test
displays the test being measured.
Here you can switch to another test from the given list. However, this
action will terminate the ongoing measurement test.
Noise Filter
displays the type of noise filter used.
Power Line Frequency
displays the frequency value of the power line.
Threshold Settings
button opens a new page that allows you to set the
VF/AC Balance
Threshold
.