TDR
202
MAX-610
Continuous
Pulse Width
is a non-editable value that is updated dependent on
whether the parameters are set to
Auto
or manual.
Dual Trace Offset
is a configurable offset parameter for the secondary
trace, with a range of -1.0 to 1.0.
Note:
If the loaded secondary trace offset = 0, then the primary and secondary
traces will overlap.
Cable Setup
button opens a new page that allows you to set the
Cable
Parameters
for the
Current Test
.
Select Graph
button changes to
Graph Selected
when pressed and
opens a new selection of function keys which allow you to maneuver
the graph as follows:
Cursor
- press the left/right arrow keys to decrease/increase the
values by moving the cursor. Press and hold the left/right arrow
keys to accelerate the cursor. Press
to recall the cursor and
place it in the middle of the current x-axis limits.
Zoom
references the middle of the x/y axis and
not
the cursor
position. Press the arrow keys as follows:
Right = x-axis zoom in.
Left = x-axis zoom out.
Up = y-axis zoom in.
Down = y-axis zoom out.
Pan
- use left/right arrow keys to pan the x-axis and the up/down
arrow keys to pan the y-axis.
1:1
resets the x-axis and y-axis limits to full scale.
Return To Test
removes these function keys and displays the
previous ones.
Result
displays the distance to the most significant reflection in text
and graphical measurements.