Safety Information
14
OTH-7000
Electrical Safety Information
Equipment Ratings
Temperature
³
Operation
³
unit connected to DC power: 0 °C to 45 °C
(32 °F to 113 °F)
³
unit connected to AC power (with AC/DC power
adapter): 0 °C to 40 °C (32 °F to 104 °F)
³
Storage
³
–40 °C to 70 °C (–40 °F to 158 °F)
Relative humidity
a
a.
Measured in 0 °C to 31 °C (32 °F to 87.8 °F) range, decreasing linearly to 50 % at 40 °C (104 °F).
0 % to 95 % non-condensing
Maximum operation altitude
³
unit connected to DC power: 3000 m (9843 ft)
³
unit connected to AC power (with AC/DC power
adapter): 2000 m (6562 ft)
Pollution degree
2
Overvoltage category
I
Measurement category
Not rated for measurement categories II, III, or IV
Input power
b
b.
Not exceeding ± 10 % of the nominal voltage.
³
unit: –48 V
;2 A
c
³
AC/DC power adapter (units connected to AC power
only): 100 - 240 V
; 50/60 Hz; 1 - 0.5 A
c.
Range: –38.5 - –70 V.
C
AUTION
The use of voltages higher than those indicated on the label affixed
to your unit may damage the unit.