background image

User’s Guide 
Mini Analog 
MultiMeter 

Model 38073 

Introduction 

Congratulations on your purchase of the Extech model 38073 Analog Multimeter. 
This device measures AC/DC Voltage, DC Current, Resistance, dB, and has battery 
test, displaying measurements in analog format. Properly used, this meter will 
provide many years of reliable service. 

Safety 

International Safety Symbols 

This symbol, adjacent to another symbol or terminal, indicates the user 
must refer to the manual for further information. 
 
This symbol, adjacent to a terminal, indicates that, under normal use, 
hazardous voltages may be present 
 
Double insulation 

WARNINGS 

1.  When measuring voltage, ensure that the function switch is not in the current or 

resistance positions. Before taking any measurement ensure that the function 
switch is set to the appropriate position. 

2.  Use great care taking measurements when the voltages are greater than 25VAC 

rms or 35VDC. These voltages are considered a shock hazard. 

3.  When taking current measurements ensure that the circuit under test is not "live". 

Remove power to the circuit, apply the test leads, and then power the circuit. 

4.  For resistance measurements, ensure that the circuit or component is de-

energized. 

5.  Use care when measuring current transformers, high voltage may result at the 

meter's terminals if an open circuit exists. 

6.  Ensure that the test leads are in good repair before each use. Have the meter 

serviced if it appears that the test leads are damaged. 

7.  Do not exceed the measurement range limits specified by the manufacturer. 
8.  Replace fuse with fuse of the same rating only. 
9.  Before opening the case to replace battery or fuse, disconnect the test leads from 

any circuit or component under test and turn the function switch to the OFF 
position. 

 
A UL mark does not indicate that this product has been evaluated for the accuracy of 
its readings. 
 
 

Specifications 

 Range 

Accuracy 

Sensitivity 

DC Voltage 

2.5, 10, 50, 250, 300V 

± 5% of full scale 

2K

 / Volt sensitivity 

AC Voltage 

10, 50, 300V 

± 5% of full scale 

2K

 / Volt sensitivity 

DC Current 

0.5, 10, 250mA 

± 5% of full scale 

 

Resistance 

Rx10, Rx1000 (Rx1k) 

± 5% of full scale 

 

Decibels 

-10 to 56 dB 

 

(0dB = 1mW 
dissipated in 600

1.5V 

± 10% of full scale 

Load current: 125mA 

Battery Test 

9V 

± 10% of full scale 

Load current: 10mA 

Display type

 Analog 

Continuous 

Operating Temp

41 to 104

o

F (5 to 40

o

C) 

Storage Temp 

41 to 104

o

F (5 to 40

o

C) 

Battery 

One 1.5V AA battery 

Battery life

 

10 hours  (continuous) 

Fuses

 

500mA, 250V fast acting 

Dimensions/Weight

 

3.5x6.1x1.9" (88x155x49mm) / 10.2 oz. (288g) approx. 

Safety 

For indoor use and in accordance with Overvoltage 
Category II, Pollution Degree 2. Category II includes local 
level, appliance, portable equipment, etc., with transient 
overvoltages less than Overvoltage Category III. 

 
 
 

Meter Description 

1. Analog 

Display 

2. Display 

zero 

adjust 

3. 0

 adjust 

4. 

Function select switch 

5. Test 

Leads 

 
 
 
 
 

Operation

 

NOTICE

: Read and understand all 

warning

 statements listed in the safety 

section of this operation manual prior to using this meter. 

AC/DC Voltage Measurements  

1.  Set the function switch to the desired voltage range position. If the range of the 

measurement is not known beforehand, select the highest range first and then 
work down range by range. 

2.  Connect the test leads in parallel to the circuit under test. 
3.  Read the voltage measurement on the analog display. 

DC mA Current Measurements 

1.  For mA current measurements set the function switch to the desired current range 

position and insert the red test lead into the mA terminal. If the range of the 
measurement is not known beforehand, select the highest range first and work 
down range by range as needed. 

2.  Remove power from the circuit under test and open the circuit at the point where 

you wish to measure current. 

3.  Touch the black test probe tip to the negative side of the circuit and touch the red 

test probe tip to the positive side of the circuit. 

4.  Apply power to the circuit. 
5.  Read the current on the analog display. If the analog pointer fails to move during 

current measurements, check the meter’s internal fuse. 

Resistance Measurements 

1.  Set the function switch to the desired resistance range position. 
2.  Touch the test probe tips across the circuit or component under test. It is best to 

disconnect one side of the resistor under test so the rest of the circuit will not 
interfere with the resistance reading. 

3.  Read the resistance on the analog display.  

Battery Test

 

1.  Select the 1.5V or 9V BAT position using the function select switch. 
2.  Connect the red test lead to the positive side of the 1.5V or 9V battery and the 

black test lead to the negative side of the 1.5V or 9V battery. 

3.  Read the analog display. The red display region indicates a weak (bad) battery, 

the green display region indicates a good battery, and the ? region indicates that 
the battery is on the verge of weakening. 

Decibel (dB) Measurements

 

1.  Set the function switch to the desired AC Voltage position. 
2.  Connect the test leads across the circuit under test. 
3.  For the 10V AC range, read dB directly from the analog scale. 
4.  For other ranges, calculate the measurement using the table below: 
5.  Circuits with dc component, require a capacitor between test leads and circuit. 

dB Range 

-10 to 22dB 

4 to 36 dB 

24 to 56 dB 

AC Range 

10V 

50V 

500V 

Add value to reading 

14 

34 

Maintenance 

WARNING: 

To avoid electrical shock, disconnect the meter from any circuit, remove 

the test leads from the input terminals and turn OFF the meter before opening the 
case. Do not operate with open case. Install only the same type of fuse or equivalent 

Cleaning and Storage 

Periodically wipe the case with a damp cloth and mild detergent; do not use 
abrasives or solvents. If the meter is not to be used for periods of longer than 60 
days, remove the battery and store it separately 

Battery and Fuse replacement 

1.  Remove the meter's protective holster 
2.  Remove the three rear Phillips head screws and open the meter case. 
3.  Replace the AA battery and/or fuse (500mA @250V, fast acting) 
4. Re-assemble the meter 

Warranty 

EXTECH INSTRUMENTS CORPORATION warrants this instrument to be free of 
defects in parts and workmanship for one year from date of shipment (a six month 
limited warranty applies on sensors and cables). If it should become necessary to 
return the instrument for service during or beyond the warranty period, contact the 
Customer Service Department at (781) 890-7440 ext. 210 for authorization. 

A Return 

Authorization (RA) number must be issued before any product is returned to 
Extech.

 The sender is responsible for shipping charges, freight, insurance and 

proper packaging to prevent damage in transit. This warranty does not apply to 
defects resulting from action of the user such as misuse, improper wiring, operation 
outside of specification, improper maintenance or repair, or unauthorized 
modification. Extech specifically disclaims any implied warranties or merchantability 
or fitness for a specific purpose and will not be liable for any direct, indirect, incidental 
or consequential damages. Extech's total liability is limited to repair or replacement of 
the product. The warranty set forth above is inclusive and no other warranty, whether 
written or oral, is expressed or implied.

 

Repair and Calibration Services 

Extech offers complete repair and calibration services

 for all Extech products. 

For periodic calibration, NIST certification or repair of any Extech product, call 
customer service for details on services available. Extech recommends that 
calibration be performed on an annual basis to insure calibration integrity. 
 

Copyright © 2000 Extech Instruments Corporation. All rights reserved 

including the right of reproduction in whole or in part in any form. 

 

38073   V1.3     9/00 

 