380900-EN v3.1 07/13
14
Voltage and Current Supplemental Accuracy Specifications for Dual Channel use
Add the accuracy figures listed below to the accuracy figures in the tables above for
voltage and current when using the meter’s two channels simultaneously
Note:
For DC to 400Hz use only
ACV
(CH2)
DCV
(CH2)
ACA
(CH2)
DCA
(CH2)
ACV
(CH1)
±1%±(V1-V2)
(200 PPM)
±0.5%±(V1-V2) *
(20 PPM)
±0.1µA/V (µA,CH2)
±1µA/V (mA, CH2)
±0.2mA/V (A, CH2)
±2mV/A (mV, CH1)
±5mV/A (V, CH1)
±0.25%
DCV
(CH1)
±0.5%±(V1-V2) *
20 PPM
±0.25% ±0.25% ±0.25%
Legend:
V1: The voltage reading of Channel 1
V2: The voltage reading of Channel 2
PPM: parts per million
Function Range Resolution
Accuracy
Resistance
999.99
Ω
0.01
Ω
±(0.19%+8dgts)
9.9999K
Ω
0.0001K
Ω
±(0.09%+3dgts)
99.999K
Ω
0.001K
Ω
±(0.09%+3dgts)
999.99K
Ω
0.01K
Ω
±(0.09%+3dgts)
9.9999M
Ω
0.0001M
Ω
±(0.2%+6dgts)
40.000M
Ω
0.001M
Ω
±(1%+6dgts)
Capacitance
(auto range)
60.00nF 0.01nF
±(0.8%+5dgts)
600.0nF 0.1nF
±(1.5%+5dgts)
6.000µF 0.001µF
±(1.5%+5dgts)
60.00µF 0.01µF
±(2.0%+5dgts)
490.0µF 0.1µF
±(3.5%+5dgts)
Capacitance
(manual range)
99.99nF 0.01nF
±(0.8%+5dgts)
999.9nF 0.1nF
±(1.5%+5dgts)
9.999µF 1µF
±(1.5%+5dgts)
99.99µF 0.01µF
±(2.0%+5dgts)
999.9µF 0.1µF
±(3.5%+5dgts)
Frequency
(TTL)
1.000Hz-2MHz 0.01Hz-0.0001MHz
±(0.005%+4dgts)
Frequency
(AC sine wave)
1.000Hz-200KHz 0.01Hz-100Hz
±(0.02%+4dgts)
Sine wave sensitivity: 100mV@mv range, 1V@V range
Duty Cycle
(%, TTL, 1Hz to
600KHz)
0.001%-9.999% 0.001%
±(30d/KHz+30dgts)
10.00%-100.00% 0.01%
±(3d/KHz+3dgts)
Duty Cycle
(%, sine wave)
0.001%-100.00% 0.001%-0.01%
Not
Specified