5. Using the Digital Multimeter
Model 381395 Version 1.0 January 2008
5-2
5.1.3. COMPARE MODE
c
Æ
COMPARE MODE
Æ
d
Press
to open Setup
e
Press
to change Edit mode
PATH:
Æ
Æ
Selects “COMPARE MODE”
Æ
In Compare mode, the meter determines if a measurement is within, above, or below a
designated range.
In this mode, the meter compares the input value to a high and low limit value. A -LO- is
displayed if the input value is lower than the low limit, PASS if the input value is within the
high and low limit values and -HI- if the input is greater than the high limit.
5.1.4. TREND PLOT
c
Æ
TREND MODE
Æ
d
Press
to open Setup
e
Press
to change Edit mode
PATH:
Æ
Æ
Selects “TREND PLOT”
Æ
Displays the trend graph with the reference value as the center value.
[F1] SetUp: The Reference value, horizontal and vertical axis values (REF, H/Div, V/Div)
are displayed. The REF is the centerline value on the trend graph, H/Div is the X axis (time)
and V/DIV is the Y axis (amplitude).
[F3] Clear: Erase the graph on display.
[F4] Start: Starts the plotting using the REF value as the centerline.