Harmonics of AC Voltage in Percentage
(1 to 99
th
order, minimum voltage at 50 or 60 Hz > AC 80V. If
the voltage is 0 at 50 or 60 Hz, all the percentage displays will be 0.)
Range Resolution
Accuracy
1 to 20
th
0.1%
±2%
20 to 49
th
0.1%
4% of reading ±2.0%
50 to 99
th
0.1%
6% of reading ±2.0%
Harmonics of AC Voltage in Magnitude
(1 to 99
th
order, minimum voltage at the 50 or 60 Hz > AC 80V,
VT=1)
Range Resolution
Accuracy
1 to 20
th
±2% ± 0.5V
20 to 49
th
4% of reading ±0.5V
50 to 99
th
0.1V
6% of reading ±0.5V
Harmonics of AC Current in Percentage
(1 to 99
th
order, minimum current at the 50 or 60 Hz > 20 A. If the
current is 0 at 50 or 60 Hz, all the percentage (%) display is 0.)
1000A Clamp
Range Resolution
Accuracy
1 to 20
th
0.1%
±2%
20 to 49
th
0.1%
4% of reading ±2.0%
50 to 99
th
0.1%
6% of reading ±2.0%
Harmonics of AC Current in Magnitude
(1 to 99
th
order, minimum current at the 50 or 60 Hz > 20A,
CT=1)
1000A Clamp
Range Resolution
Accuracy
1 to 20
th
±2% of reading ±0.4A
20 to 49
th
±4% of reading ±0.4A
50 to 99
th
0.1A
±6% of reading ±0.4A
Harmonic of AC Current in Percentage
(1 to 99
th
order, minimum current at the 50 or 60 Hz > 20 A. If the
current is 0 at 50 or 60 Hz, all the percentage (%) display is 0.)
100A Clamp
Range Resolution
Accuracy
1 to 10
th
0.1 %
±0.2% of reading ±1%
11 to 20
th
0.1 %
±2% of reading ±1%
21 to 50
th
(A range)
0.1 %
±5% of reading ±1%
21 to 50
th
(mA range)
0.1 %
±10% of reading ±1%
51 to 99
th
0.1 %
±35% of reading ±1%