382095 V4.1 7/06
36
Harmonics of AC Current in Magnitude
(1 to 99
th
order, minimum current at the 50 or 60 Hz > 20A,
CT=1)
100A Clamp
Range Resolution
Accuracy
1 to 10
th
±0.2% of reading ±7dgts
11 to 20
th
±2% of reading ±7dgts
21 to 50
th
(A range)
±5% of reading ±7dgts
21 to 50
th
(mA range)
±10% of reading ±7dgts
51 to 99
th
0.1mA / 0.1A
±35% of reading ±7dgts
Harmonic of AC Current in Percentage
(1 to 99
th
order, minimum current at 50 or 60 Hz is greater than
20A. If the current is 0 at 50 or 60 Hz, all the percentage (%) display is 0)
3000A Flexible Clamp
Range Resolution
Accuracy
1 to 20
th
0.1
%
±2%
21 to 50
th
0.1
%
±6%
51 to 99
th
0.1
%
±10%
Harmonics of AC Current in Magnitude
(1 to 99
th
order, minimum current at the 50 or 60 Hz > 20A,
CT=1)
3000A Flexible Clamp
Range Resolution
Accuracy
1 to 20
th
±2% of range ±0.4A
21 to 50
th
±4% of range ±0.4A
51 to 99
th
0.1A
±6% of range ±0.4A
Harmonics of AC Current in Magnitude
(1 to 99
th
order, minimum current at 50 or 60 Hz is greater than
300A)
3000A Flexible Clamp
Range Resolution
Accuracy
1 to 20
th
±2% of range ±40A
21 to 50
th
±4% of range ±40A
51 to 99
th
0.1A
±6% of range ±40A
Power Factor
(PF)
Range Resolution
Accuracy
0.00 to 1.00
0.01
± 0.04 (>20V and >30A)