background image

407113-EN v3.6   4/14 

6

Air Flow (Volume) Measurements (CMM / CFM) 

1.  Connect the sensor to the sensor input jack on top of the meter. 

2.  Turn on the meter using the POWER button. 

3.  Select the FLOW mode using the FLOW/VELOCITY button. The LCD will display 

FLOW CFM

 or 

FLOW CMM

 when the flow function has been selected. 

4.  Select the desired air flow units: CMM (cubic meters per minute) or CFM (cubic feet per minute) 

using the UNIT button. The LCD will reflect the selection. 

5.  Measure the dimensions of the duct or vent and calculate the area in square feet or square 

meters

Note:

 If the dimensional measurements are made in inches (or cm), convert them feet (or meters) 

before calculating the square area. 

6.  Press the AREA button to begin entering the area in m

2

 or ft

2

. Use the 

 button to increment 

the flashing digit, use the

 button to decrement, use the  button to select the next digit, and 

use the   button to set the decimal point.  

7.  Press the ENTER/RESET button after the area value has been entered. The bottom display will 

indicate the area entered in ft

2

 or m

2

. The main LCD displays the Air Flow in CFM (cubic feet per 

minute) or CMM (cubic meters per minute). If the CFM or CMM reading exceeds 9999, use the 
displayed X10 or X100 multiplier to calculate the reading. 

8.  Note that for FLOW measurements, three modes apply: The normal, default mode, where the 

actual flow is indicated and the two modes described below under 

2/3 MAX Flow

 and 

AVG 

Flow

.

Note:

 The temperature function is not active in the FLOW mode. 

2/3V MAX Flow Mode 

In this mode the meter will display two-thirds the measured Flow (Volume). To access this mode, 
ensure that the meter is in the FLOW mode (via the FLOW/VELOCITY button) and then press the 
FLOW MODE button until the 2/3V MAX display icon appears. Now the Flow display will be 2/3 the 
actual measurement. To return to the normal display mode, press the FLOW MODE button until the 
2/3VMAX and the AVG icons switch off. 

AVG (Average) Flow Mode 

In this mode the meter will display the AVERAGE Flow for up to 20 readings. To access this mode, 
ensure that the meter is in the FLOW mode (via the FLOW/VELOCITY button) and press the FLOW 
MODE button until the AVG display icon appears. Now each time the AVG/START button is pressed 
a reading will be taken and averaged. The lower LCD display becomes a 20-reading counter and 
with each reading taken the counter is incremented. Allow 3 or 4 seconds between AVG/START 
button presses for the display counter to update. The main LCD displays the averaged Air Flow 
value. To return to the normal display mode, press the FLOW MODE button until the AVG icon 
switches off. 

RS-232 PC Interface 

The 407113 is equipped with a 3.5mm phone jack (meter top) for connection to a PC for data 
acquisition purposes. For streaming of data to a PC via the RS232 Output jack, the optional 
407001-USB kit (RS232 to USB cable and driver CD) along with the 407001 software  
at  are required. Instructions for use are provided with the data acquisition 
software/hardware kits. 

www.

GlobalTestSupply

.com

Find Quality Products Online at:

sales@GlobalTestSupply.com

Summary of Contents for 407113

Page 1: ...User Guide Heavy Duty CFM Thermo Anemometer Model 407113 www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com ...

Page 2: ...n and CFM ft 3 min Temperature C and F Data Hold Freezes displayed reading Sampling rate One 1 reading per second Sensors Air velocity flow sensor Metal angled vane arms with low friction ball bearing Temp Sensor Precision thermistor MIN MAX Records Recalls lowest and highest readings Auto Power OFF Automatic shut off after 15 minutes PC Interface RS 232 Serial Communications 16 bit data stream Ov...

Page 3: ...iles per hour 1 1 78 3 mph 0 1 mph 2 0 4m h knots nautical miles per hour 1 0 to 68 knots 0 1 knots 2 0 4knots Air Flow Measurements Range Resolution Area CMM cubic meters per minute 0 999 900 m 3 min 0 001 to 100 0 to 9 999m 2 CFM cubic feet per minute 0 999 900 ft 3 min 0 001 to 100 0 to 9 999ft 2 Air Temperature Range Resolution Accuracy 32 to 175 o F 0 to 80 o C 0 1 o F C 1 5 o F 0 8 o C www G...

Page 4: ...nterface jack 3 LCD Display 4 Vane 5 Rubber Holster 6 Keypad 7 Battery compartment rear Note To access the rear battery compartment first remove the rubber holster that jackets the meter www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com ...

Page 5: ...when the display is in Data Hold mode 3 Press HOLD again to return to normal operation Maximum and Minimum Recording The 407113 allows the user to record and view the highest MAX and lowest MIN readings 1 Press the RECORD RECALL button once The REC indicator will appear on the display and the meter will begin recording the MAX and MIN values 2 Press the RECORD RECALL button to stop the MAX MIN rec...

Page 6: ...e The temperature function is not active in the FLOW mode 2 3V MAX Flow Mode In this mode the meter will display two thirds the measured Flow Volume To access this mode ensure that the meter is in the FLOW mode via the FLOW VELOCITY button and then press the FLOW MODE button until the 2 3V MAX display icon appears Now the Flow display will be 2 3 the actual measurement To return to the normal disp...

Page 7: ...ations NOTE Measurements made in inches must be converted to feet or meters before using the above formulae Unit of Measure Conversion Table m s ft min knots km h MPH 1 m s 1 196 87 1 944 3 6 2 24 1 ft min 0 00508 1 0 00987 0 01829 0 01138 1 knot 0 5144 101 27 1 1 8519 1 1523 1 km h 0 2778 54 69 0 54 1 0 6222 1 MPH 0 4464 87 89 0 8679 1 6071 1 Width W Height H Area A Width W x Height H CFM ft3 min...

Page 8: ...r any direct indirect incidental or consequential damages FLIR s total liability is limited to repair or replacement of the product The warranty set forth above is inclusive and no other warranty whether written or oral is expressed or implied Calibration Repair and Customer Care Services FLIR Systems Inc offers repair and calibration services for the Extech Instruments products we sell NIST certi...

Page 9: ... conséquent direct indirect ou accidentel La responsabilité totale de FLIR est limitée à la réparation ou au remplacement du produit La garantie définie ci dessus est inclusive et aucune autre garantie écrite ou orale n est exprimée ou implicite Calibrage réparation et services après vente FLIR Systems Inc offre des services de calibrage et de réparation pour les produits Extech Instruments que no...

Page 10: ...sito determinado y no será responsable por cualesquier daños directos indirectos incidentales o consecuentes La responsabilidad total de FLIR está limitada a la reparación o reemplazo del producto La garantía precedente es inclusiva y no hay otra garantía ya sea escrita u oral expresa o implícita Servicios de calibración reparación y atención a clientes FLIR Systems Inc ofrece servicios de reparac...

Reviews: