412355A-EU-EN v3.3 07/13
4
MEM BUTTON
The memory feature provides 5 user settable source values for stepped calibration outputs. The
feature is available for Voltage, mA and % in the SOURCE mode. The memory values are stored in
nonvolatile memory and are not erased when power is turned off.
Sourcing from the stored memory values:
1.
Select the Source mode
2.
Press the MEM button. The M1 icon (memory location 1) will appear in the display and the
value stored in that location will be displayed and sourced.
3.
Repeated pressing of the MEM button will step through the 5 memory locations.
NOTE
: The “
SOURCE”
icon will blink when the output value has not reached a stable level.
The common cause for the “
SOURCE”
icon to remain blinking is that the load impedance is
too high in the current mode or too low in the voltage mode.
Storing values into memory:
1.
Select and display a memory location (M1 to M5)
2. Press
the
▲
▼
buttons to adjust the display to the desired source value.
3.
Press and hold the MEM button for > 2 seconds. The displayed value will be stored into the
displayed memory location.
Default memory values.
Five common source values are permanently programmed in memory as the default values.
These values can be replaced by user selected values.
To reset the meter to the default memory values:
1.
Turn the meter on and select the SOURCE mode.
2.
Press and hold the POWER button for >4 seconds. will briefly appear in the display
and the default values will be stored into memory.
Default Memory Values
M1
M2 M3 M4 M5
mV 0mV
500mV
1000mV
1500mV
2000mV
V 2V 5V 10V 15V 20V
mA 4mA 8mA 12mA 16mA 20mA
% 0% 25% 50% 75% 100%
Over-range / Under-range Indication
Signals above or below the units ranges will be indicated by “HHHH” for above and “LLLL” for
below.