background image

Model 421501 Version 1.3  February 2006  

Warranty 

EXTECH INSTRUMENTS CORPORATION

 warrants this instrument to be free of defects in parts and 

workmanship for 

one year

 from date of shipment (a six month limited warranty applies to sensors and cables). 

If it should become necessary to return the instrument for service during or beyond the warranty period, 
contact the Customer Service Department at (781) 890-7440 ext. 210 for authorization or visit our website 
www.extech.com for contact information. A Return Authorization (RA) number must be issued before any 
product is returned to Extech. The sender is responsible for shipping charges, freight, insurance and proper 
packaging to prevent damage in transit. This warranty does not apply to defects resulting from action of the 
user such as misuse, improper wiring, operation outside of specification, improper maintenance or repair, or 
unauthorized modification. Extech specifically disclaims any implied warranties or merchantability or fitness for 
a specific purpose and will not be liable for any direct, indirect, incidental or consequential damages. Extech's 
total liability is limited to repair or replacement of the product. The warranty set forth above is inclusive and no 
other warranty, whether written or oral, is expressed or implied. 
 

Specifications 

Display 

 

3-1/2 digit LCD (2000 counts) 

 

Battery Power Supply

 

9V DC battery.

 

Battery life

 

200 hours typical

 

Dimensions / Weight 

 

7.5 x 3.6 x 2.1” (192 x 91 x 52.5mm); 11.7oz. (365g)

 

Range / Resolution

 

-58 to 2000

o

F (-50 to 1300

o

C); 1

o

 or 0.1

o

F/

o

Important Note

: The supplied thermocouple can only 

be used to measure temperature up to 500

 o

F (260

 o

C). 

To measure higher temperatures, a high temperature 
Type K thermocouple must be used.

 

Temperature units

 

Selectable 

o

C or 

o

F temperature units 

 

±(0.3% of r 1

o

C) from -50

o

C to 1000

o

C

 

±(0.5% of r 1

o

C) from 1000

o

C to 1300

o

C

 

Accuracy of meter

 

±(0.3% of r 2

o

F) from -58

o

F to 2000

o

F

 

Temperature Coefficient

 

0.1 times the applicable accuracy specifications per 

o

from 0

o

C to 18

o

C and 28

o

C to 50

o

C (32

o

F to 64

o

F and 

82

o

F to 122

o

F)

 

Water resistant housing 

Gasket protected front panel 

Input Protection

 

24VAC/DC max. input on any combination of inputs

 

Reading rate

 

2.5 times per second

 

Thermocouple connection

 

Meter accepts standard subminiature connector

 

Supplied thermocouple

 

Teflon insulated Type K thermocouple (4’ length) 

Max temperature: 500

o

F (260

o

C); For higher 

temperatures, a high temp thermocouple must be used. 

Accuracy: ±4

o

F (±2.2

o

C) or +0.75% of rdg 

(whichever is greater) from 32 to 1472

o

F (0 to 800

o

C)

 

Operating conditions

 

32 to 122

o

F (0 to 50

o

C); less than 80% RH

 

Storage conditions

 

-4 to 140

o

F (-20 to 60

o

C); less than 70% RH

 

Summary of Contents for 421501

Page 1: ... meter measures temperature with a Type K thermocouple The thermocouple attaches to the meter with a sub miniature plug The meter s display is backlit and includes a Data Hold feature Other features include Max Temperature and selectable units resolution This meter is shipped fully tested and calibrated and with proper use will provide years of reliable service 0 1 1 ...

Page 2: ...n or oral is expressed or implied Specifications Display 3 1 2 digit LCD 2000 counts Battery Power Supply 9V DC battery Battery life 200 hours typical Dimensions Weight 7 5 x 3 6 x 2 1 192 x 91 x 52 5mm 11 7oz 365g Range Resolution 58 to 2000 o F 50 to 1300 o C 1 o or 0 1 o F o C Important Note The supplied thermocouple can only be used to measure temperature up to 500 o F 260 o C To measure highe...

Page 3: ...ching an object or in a liquid solution Read the temperature measurement on the LCD Important Note The supplied thermocouple can only be used to measure temperature up to 500 o F 260 o C To measure higher temperatures a high temperature Type K thermocouple must be used Selecting o C o F units of measure Press the o C o F key to select the desired temperature unit of measure Selecting the Resolutio...

Page 4: ...rews that secure the battery compartment 3 Open the compartment and replace the 9V battery 4 Secure the meter s battery compartment and rubber jacket before using the meter WARNING To avoid possible electrical shock disconnect the thermocouple before removing the battery cover Cleaning Use a damp cloth with a mild detergent to periodically wipe the meter housing and display area Do not use abrasiv...