background image

Model 42525A Version 4.0   5/05 

Operating Instructions 

Infrared Measurement Procedure 

1.  Turn the meter ON by pressing the POWER button. The meter display will countdown 

(99999, 88888, 77777…00000) during the self-test procedure. When the self-test is 
complete, the display will indicate a temperature reading, the unit of measure, the 
emissivity setting, and the measurement type (IR for non-contact or Type K for contact 
measurements). 

2.  Use the IR / TYPE K button to set the display to the IR mode (IR will appear in the 

lower left-hand side of the LCD). 

3.  Set the temperature units (

o

C or 

o

F) using the C/F button. 

4.  Point the IR sensor toward the object under test (to assist with targeting, press the 

meter trigger to activate the laser pointer). 

5.  This meter has a “distance to spot ratio” of 6:1. Use the diagram below to determine 

the proper distance to hold the meter from an object. For example, if the object under 
test has a diameter of 4”, hold the meter at a distance of approximately 26”. 

6.  Read the LCD display for the temperature measurement. Note that the current 

emissivity setting is shown on the lower right-hand corner of the LCD. 

 

 

 

 

 

 

 

 

 

 

Infrared Measurement Considerations 

1.  The 42525A automatically compensates for ambient temperature deviations; however 

it may take up to 30 minutes to adjust to extremely wide ambient temperature changes. 

2.  When low temperature measurements are taken followed by high temperature 

measurements, several minutes are required for stabilization before the high 
temperature measurements can be made accurately. 

3.  Measurement Field / Distance: For optimum accuracy, the object should be 1.5 to 2 

times larger than the spot size shown in the diagram above and on the meter side. 

4.  Measurement Interference: Objects having low emissivity or objects with low 

temperature yet high emissivity emit little IR energy. Such objects are adversely 
affected by IR energy radiated from nearby objects having high emissivity and 
temperature. For example, when such objects are measured in sunlight, erratic 
readings occur because of the powerful radiation (sunlight) reflected off of the object's 
surface into the 42525A sensor. 

5.  If the surface of the object under test is covered with frost, oil, grime, etc. clean before 

taking measurements. 

6.  If an object's surface is highly reflective, apply masking tape or flat black paint 

(emissivity 0.95) before measuring. 

150

125

100

75

50

mm

6

5

4

3

2

inch

1000

840

670

500

330

mm

40

33

26.4

20

13

inch

Diameter of Spot

Distance to Object

Summary of Contents for 42525A

Page 1: ...User s Manual InfraRed Thermometer with Laser Pointer MODEL 42525A ...

Page 2: ...orkmanship for one year from date of shipment a six month limited warranty applies to sensors and cables If it should become necessary to return the instrument for service during or beyond the warranty period contact the Customer Service Department at 781 890 7440 ext 210 for authorization or visit our website www extech com for contact information A Return Authorization RA number must be issued b...

Page 3: ...Resolution 4 to 752o F 20 to 400o C 0 1o C F Accuracy 3 of reading or 6o F 3o C whichever is greater Accuracy notes 1 Accuracy specified for readings between 10 and 350o C 2 Accuracy specified for ambient temperature 23o C 5o C 3 Accuracy specified for an emissivity of 0 95 4 Accuracy tested using 2 25 diameter black body at a distance of 6 5 Testing environment RF field strength 3 V M and frequen...

Page 4: ...also the DOWN button Used to access the RELATIVE mode of operation Also used as the DOWN arrow button for setting the emissivity Record button Used when viewing the MAX and MIN readings Data Hold button Used to freeze the displayed reading 3 POWER button Turns meter ON or OFF 4 IR Type K button Selects IR non contact or Type K contact operation 5 RS 232 PC Interface jack 6 IR Sensor 7 Laser pointe...

Page 5: ...Measurement Considerations 1 The 42525A automatically compensates for ambient temperature deviations however it may take up to 30 minutes to adjust to extremely wide ambient temperature changes 2 When low temperature measurements are taken followed by high temperature measurements several minutes are required for stabilization before the high temperature measurements can be made accurately 3 Measu...

Page 6: ... emissivity 3 Press and hold the Emissivity button until the display stops flashing to exit Emissivity Factors for Common Materials Material under test Emissivity Material under test Emissivity Asphalt 0 90 to 0 98 Cloth black 0 98 Concrete 0 94 Skin human 0 98 Cement 0 96 Lather 0 75 to 0 80 Sand 0 90 Charcoal powder 0 96 Soil 0 92 to 0 96 Lacquer 0 80 to 0 95 Water 0 92 to 0 96 Lacquer matt 0 97...

Page 7: ...e held down correctly While continuing to hold down the two keys press the UP or DOWN buttons to set the offset value on the larger digits Release all keys to return to normal operation Type K Thermocouple Measurement Procedure 1 Connect a Type K Thermocouple to the temperature input jack on the meter 2 Power the meter by pressing the POWER button 3 Use the IR TYPE K button to select Type K mode d...

Page 8: ...he battery compartment cover Cleaning Wipe meter with a damp cloth when necessary A mild detergent may be used but avoid solvents abrasives and harsh cleansers Do not allow moisture inside the meter Storage Remove the battery if the meter is to be stored for long periods Do not store the meter in an environment that exceeds the published operating temperature and humidity specifications Calibratio...

Reviews: