Model 42525A Version 4.0   5/05 

Max/Min Recording 

The Max/Min Recording feature offers the user to view only the highest (maximum MAX) 
or the lowest (minimum MIN) reading. Press the REC button once to activate this feature 
(the REC symbol will appear on the LCD). Press the REC button again to view only the 
maximum reading (MAX). Press the REC button again to view only the minimum reading 
(MIN). Press and hold the RECORD button again until the REC/MAX/MIN icons switch off 
to return to normal operation.

 

Laser Pointer Operation 

This 42525A is equipped with a laser pointer that is used to aim the 
IR sensor more accurately. Simply press the meter trigger to activate 
the laser pointer. Be extremely cautious when directing the laser 
beam, shining the laser beam into someone's eye can cause serious 
injury. Point the laser beam toward the device under test whose 
temperature is to be measured. 

Offset Adjustment 

The OFFSET function allows the user to adjust the temperature readings in fine detail to 
match an external calibrator, to compensate for display drift, or for other diagnostic uses. 
Press and hold the HOLD and the REC buttons simultaneously. The small digits on the 
lower right will match the temperature display of the larger digits when the two buttons are 
held down correctly. While continuing to hold down the two keys, press the UP or DOWN 
buttons to set the offset value on the larger digits. Release all keys to return to normal 
operation.  

Type K Thermocouple Measurement Procedure 

1.  Connect a Type K Thermocouple to the temperature input jack on the meter. 

2.  Power the meter by pressing the POWER button. 

3.  Use the IR / TYPE K button to select Type K mode (display will show ‘K’). 

4.  Set the temperature units (degrees C or F) using the C/F button. 

5.  Place the Thermocouple in the area where the temperature is to be measured.  

6.  Read the LCD display for the temperature measurement. 

 

Type K Measurement Considerations 

When a Type K Thermocouple is first connected to the meter and the thermocouple and 
meter temperatures are different, several minutes should be allotted so that the 
temperature of the socket matches the temperature of the thermocouple. The cold junction 
compensation circuitry can then operate efficiently, allowing the meter to provide 
measurements with optimum accuracy. 

Auto Power Off 

The meter will automatically power off after approximately 10 minutes in both IR and 
Thermocouple measurement modes. To disable this feature, press the REC button. 

 

RS-232 Serial PC Interface 

This instrument includes a built-in RS-232 serial data port. This interface is intended for 
use with the optional Data Acquisition Software and Cable . For more information, contact 
Extech or refer to the manual with software package. 

 

Summary of Contents for 42525A

Page 1: ...User s Manual InfraRed Thermometer with Laser Pointer MODEL 42525A ...

Page 2: ...orkmanship for one year from date of shipment a six month limited warranty applies to sensors and cables If it should become necessary to return the instrument for service during or beyond the warranty period contact the Customer Service Department at 781 890 7440 ext 210 for authorization or visit our website www extech com for contact information A Return Authorization RA number must be issued b...

Page 3: ...Resolution 4 to 752o F 20 to 400o C 0 1o C F Accuracy 3 of reading or 6o F 3o C whichever is greater Accuracy notes 1 Accuracy specified for readings between 10 and 350o C 2 Accuracy specified for ambient temperature 23o C 5o C 3 Accuracy specified for an emissivity of 0 95 4 Accuracy tested using 2 25 diameter black body at a distance of 6 5 Testing environment RF field strength 3 V M and frequen...

Page 4: ...also the DOWN button Used to access the RELATIVE mode of operation Also used as the DOWN arrow button for setting the emissivity Record button Used when viewing the MAX and MIN readings Data Hold button Used to freeze the displayed reading 3 POWER button Turns meter ON or OFF 4 IR Type K button Selects IR non contact or Type K contact operation 5 RS 232 PC Interface jack 6 IR Sensor 7 Laser pointe...

Page 5: ...Measurement Considerations 1 The 42525A automatically compensates for ambient temperature deviations however it may take up to 30 minutes to adjust to extremely wide ambient temperature changes 2 When low temperature measurements are taken followed by high temperature measurements several minutes are required for stabilization before the high temperature measurements can be made accurately 3 Measu...

Page 6: ... emissivity 3 Press and hold the Emissivity button until the display stops flashing to exit Emissivity Factors for Common Materials Material under test Emissivity Material under test Emissivity Asphalt 0 90 to 0 98 Cloth black 0 98 Concrete 0 94 Skin human 0 98 Cement 0 96 Lather 0 75 to 0 80 Sand 0 90 Charcoal powder 0 96 Soil 0 92 to 0 96 Lacquer 0 80 to 0 95 Water 0 92 to 0 96 Lacquer matt 0 97...

Page 7: ...e held down correctly While continuing to hold down the two keys press the UP or DOWN buttons to set the offset value on the larger digits Release all keys to return to normal operation Type K Thermocouple Measurement Procedure 1 Connect a Type K Thermocouple to the temperature input jack on the meter 2 Power the meter by pressing the POWER button 3 Use the IR TYPE K button to select Type K mode d...

Page 8: ...he battery compartment cover Cleaning Wipe meter with a damp cloth when necessary A mild detergent may be used but avoid solvents abrasives and harsh cleansers Do not allow moisture inside the meter Storage Remove the battery if the meter is to be stored for long periods Do not store the meter in an environment that exceeds the published operating temperature and humidity specifications Calibratio...