Operating Instructions 

Basic IR Measurements 

1.  Hold the meter by its handle and point it toward the surface to be measured. 

2.  Pull and hold the trigger to turn the meter on and begin testing. The temperature 

reading, the flashing ‘SCAN’ icon, the emissivity, and the unit of measure will appear. 
Note: Replace the 9V battery if the display does not switch on. 

3.  Release the Trigger and the reading will hold for approximately 7 seconds (HOLD will 

appear on the LCD) after which the meter will automatically shut off. The only 
exception to this is if the TEST LOCK switch is set to ON. 

Note

: Select the temperature units (ºF/ºC) using the top switch inside the battery 

compartment 

 

 

Backlight/Laser Pointer 

1.  While pulling the Trigger, push the backlight/laser button          once to turn on the 

backlight.

 

2.  Press it again to turn on the laser pointer. When the laser is ON the laser icon     will 

appear in the LCD.

 

3.  Press the laser button to turn the backlight off.

 

4.  Pressing it again turns the laser off.

 

Note

: Backlight and Laser settings will be retained after the meter powers down. 

 

 

Over-range Indication 

If the temperature measurement exceeds the specified temperature range, the 
thermometer will display dashes in place of a temperature reading. 

 

 

Summary of Contents for 42545A

Page 1: ...User s Manual High Temperature InfraRed Thermometer with Laser Pointer MODEL 42545A ...

Page 2: ...ble for shipping charges freight insurance and proper packaging to prevent damage in transit This warranty does not apply to defects resulting from action of the user such as misuse improper wiring operation outside of specification improper maintenance or repair or unauthorized modification Extech specifically disclaims any implied warranties or merchantability or fitness for a specific purpose a...

Page 3: ...m 2 IR Sensor 3 Measurement Trigger 4 Battery and Switch Compartment 5 LCD Display 6 Push buttons 7 Handle Grip 8 Temperature Units ºC ºF Switch 9 Test Lock ON OFF Switch 10 Alarm ON OFF Switch Note There is a tripod mount on the bottom of the handle 1 2 3 4 8 9 10 ...

Page 4: ...y shut off The only exception to this is if the TEST LOCK switch is set to ON Note Select the temperature units ºF ºC using the top switch inside the battery compartment Backlight Laser Pointer 1 While pulling the Trigger push the backlight laser button once to turn on the backlight 2 Press it again to turn on the laser pointer When the laser is ON the laser icon will appear in the LCD 3 Press the...

Page 5: ...value In the AVG mode all of the readings in the current measurement session are averaged and the value is displayed HAL High Alarm setting The temperature that when exceeded causes the audible visual alarm to trip LAL Low Alarm setting The temperature that when exceeded high to low causes the audible visual alarm to trip High and Low Alarm Feature The Model 42545a has an alarm feature whereas a H...

Page 6: ...ents through transparent surfaces such as glass may not be accurate 5 Steam dust smoke etc can obscure measurements 6 The meter automatically compensates for deviations in ambient temperature However it can take up to 30 minutes for the meter to adjust to extremely wide changes 7 To find a hot spot aim the meter outside the area of interest then scan across in an up and down motion until the hot s...

Page 7: ... flat black finish For the Model 42545a the emissivity is adjustable from 0 1 to 1 00 Most organic materials and painted or oxidized surfaces have an emissivity factor of 0 95 When in doubt set the emissivity to 0 95 Emissivity Factors for Common Materials Material under test Emissivity Material under test Emissivity Asphalt 0 90 to 0 98 Cloth black 0 98 Concrete 0 94 Skin human 0 98 Cement 0 96 L...

Page 8: ...6 F 3 C 800 to 1000 F 426 to 537 C 3 of reading 9 F 4 C 1000 F 537 C Note Accuracy is specified for the following ambient temperature range 64 to 82 F 18 to 28 C Emissivity Adjustable from 0 1 to 1 00 0 95 default value Field of View D S Approx 50 1 ratio D distance S spot Laser power Less than 1mW Class II Spectral response 8 to 14 μm wavelength General Specifications Display 4 digit backlit LCD ...