background image

Model 42545

A

 V4.1   02/08 

Battery Replacement 

 
When the battery symbol     appears empty or close to empty, replace the 
meter’s 9V battery. The battery compartment is located behind the panel 
that surrounds the meter’s trigger. The panel can be pried open near the 
trigger and folded down as shown in the diagram. Replace the 9V battery 
and close the battery compartment cover. 
 
 
 
 
 
 
 

IR Measurement Notes 

1.  The object under test should be larger than the spot (target) size calculated by the field 

of view diagram (printed on the side of the meter and in this guide). 

2.  Before measuring, be sure to clean surfaces that are covered with frost, oil, grime, etc. 

3.  If an object's surface is highly reflective, apply masking tape or flat black paint to the 

surface before measuring. Allow time for the paint or tape to adjust to the temperature 
of the surface it is covering. 

4.  Measurements through transparent surfaces such as glass may not be accurate. 

5.  Steam, dust, smoke, etc. can obscure measurements. 

6.  The meter automatically compensates for deviations in ambient temperature. However, 

it can take up to 30 minutes for the meter to adjust to extremely wide changes. 

7.  To find a hot spot, aim the meter outside the area of interest then scan across (in an up 

and down motion) until the hot spot is located. 

Field of View 

The meter’s field of view is 50:1. For example, if the meter is 50 inches from the target 
(spot), the diameter of the target must be at least 1 inch. Other distances are shown in the 
field of view diagram. Note that measurements should normally be made as close as 
possible to the device under test. The meter can measure from moderate distances but the 
measurement may be affected by external sources of light. In addition, the spot size may 
be so large that it encompasses surface areas not intended to be measured. 

25”

0.64m

36”

0.91m

50”

1.27m

100”

2.54m

0.5”

1.27cm

0.72”

1.83cm

1.0”

2.54cm

2.0”

5.08cm

Distance to object

Diameter of object

Summary of Contents for 42545A

Page 1: ...User s Manual High Temperature InfraRed Thermometer with Laser Pointer MODEL 42545A ...

Page 2: ...ble for shipping charges freight insurance and proper packaging to prevent damage in transit This warranty does not apply to defects resulting from action of the user such as misuse improper wiring operation outside of specification improper maintenance or repair or unauthorized modification Extech specifically disclaims any implied warranties or merchantability or fitness for a specific purpose a...

Page 3: ...m 2 IR Sensor 3 Measurement Trigger 4 Battery and Switch Compartment 5 LCD Display 6 Push buttons 7 Handle Grip 8 Temperature Units ºC ºF Switch 9 Test Lock ON OFF Switch 10 Alarm ON OFF Switch Note There is a tripod mount on the bottom of the handle 1 2 3 4 8 9 10 ...

Page 4: ...y shut off The only exception to this is if the TEST LOCK switch is set to ON Note Select the temperature units ºF ºC using the top switch inside the battery compartment Backlight Laser Pointer 1 While pulling the Trigger push the backlight laser button once to turn on the backlight 2 Press it again to turn on the laser pointer When the laser is ON the laser icon will appear in the LCD 3 Press the...

Page 5: ...value In the AVG mode all of the readings in the current measurement session are averaged and the value is displayed HAL High Alarm setting The temperature that when exceeded causes the audible visual alarm to trip LAL Low Alarm setting The temperature that when exceeded high to low causes the audible visual alarm to trip High and Low Alarm Feature The Model 42545a has an alarm feature whereas a H...

Page 6: ...ents through transparent surfaces such as glass may not be accurate 5 Steam dust smoke etc can obscure measurements 6 The meter automatically compensates for deviations in ambient temperature However it can take up to 30 minutes for the meter to adjust to extremely wide changes 7 To find a hot spot aim the meter outside the area of interest then scan across in an up and down motion until the hot s...

Page 7: ... flat black finish For the Model 42545a the emissivity is adjustable from 0 1 to 1 00 Most organic materials and painted or oxidized surfaces have an emissivity factor of 0 95 When in doubt set the emissivity to 0 95 Emissivity Factors for Common Materials Material under test Emissivity Material under test Emissivity Asphalt 0 90 to 0 98 Cloth black 0 98 Concrete 0 94 Skin human 0 98 Cement 0 96 L...

Page 8: ...6 F 3 C 800 to 1000 F 426 to 537 C 3 of reading 9 F 4 C 1000 F 537 C Note Accuracy is specified for the following ambient temperature range 64 to 82 F 18 to 28 C Emissivity Adjustable from 0 1 to 1 00 0 95 default value Field of View D S Approx 50 1 ratio D distance S spot Laser power Less than 1mW Class II Spectral response 8 to 14 μm wavelength General Specifications Display 4 digit backlit LCD ...