AUT500-EU-EN v1.0 5/13
8
AC VOLTAGE MEASUREMENTS
WARNING:
Risk of Electrocution. The probe tips may not be long enough to contact the live parts
inside some 240V outlets for appliances because the contacts are recessed deep in the outlets. As
a result, the reading may show 0 volts when the outlet actually has voltage on it. Make sure the
probe tips are touching the metal contacts inside the outlet before assuming that no voltage is
present.
CAUTION:
Do not measure AC voltages if a motor on the circuit is being switched ON or OFF.
Large voltage surges may occur that can damage the meter.
1. Set the function switch to the ‘V’ position.
2. Press the MODE button to indicate “
AC
” on the display.
3. Insert the black test lead banana plug into the negative
COM
jack.
Insert red test lead banana plug into the positive
V
jack.
4. Touch the black test probe tip to the neutral side of the circuit.
Touch the red test probe tip to the “hot” side of the circuit.
5. Read the voltage in the display as numerical digits and as bar-graph
representation.
DC CURRENT MEASUREMENTS
CAUTION:
Do not make current measurements on the 20A scale for longer than 30 seconds.
Exceeding 30 seconds may cause damage to the meter and/or the test leads.
1. Insert the black test lead banana plug into the negative
COM
jack.
2. For current measurements up to 400mA DC, set the function switch to the
mA
position and insert the red test lead banana plug into the
400mA
jack.
3. For current measurements up to 20A DC, set the function switch to the
20A
range and insert the red test lead banana plug into the
20A
jack.
4. Press the MODE button to indicate “
DC
” on the display.
5. Remove power from the circuit under test, then open up the circuit at the
point where you wish to measure current.
6. Touch the black test probe tip to the negative side of the circuit.
Touch the red test probe tip to the positive side of the circuit.
7. Apply power to the circuit.
8. Read the current in the display as numerical digits and as bar-graph
representation.