Programming
Menu
The
meter
can
be
configured
and
calibrated
through
simple
button
presses
in
the
programming
menu.
Press
the
LEFT
button
to
access
the
menu
and
refer
to
the
menu
‘tree’
below.
The
menu
uses
UP/DOWN,
LEFT/RIGHT
presses
for
navigation
and
selection.
In
the
table
below
the
factory
default
settings
are
in
bold
with
an
asterisk.
Each
parameter
is
explained
in
detail
in
the
subsequent
sections.
Top
level
Sub
level
1
Sub
level
2
Notes
STATISTICAL
VIEW
AVG*
Average
of
a
series
of
readings
MIN
Lowest
of
a
series
of
readings
MAX
Highest
of
a
series
of
readings
NO
Number
of
sampled
readings
SDEV
Standard
deviation
of
a
series
OPTIONS
Measure
mode
Single*
One
reading
at
a
time
Continuous
Continuous
measurements
Working
mode
Direct*
Readings
are
not
stored
in
groups
Group
1…4
Store
readings
in
groups
Probe
used
Auto*
Meter
automatically
selects
mode
Fe
Ferrous
measurement
mode
No
Fe
Non
‐
ferrous
measurement
mode
Unit
settings
µ
m*
Micrometers
mils
Mils
=
mm
*
2.54
/
100
mm
Millimeters
Backlight
ON*
Enables
backlight
operation
OFF
Disables
backlight
operation
LCD
Statistics
AVG*
Average
of
a
series
of
readings
MAX
Highest
of
a
series
of
readings
MIN
Lowest
of
a
series
of
readings
SDEV
Standard
deviation
of
a
series
Auto
Power
OFF
Enable*
Allows
auto
power
off
to
operate
Disable
Defeats
the
auto
power
off
feature
LIMIT
Limit
settings
High
Limit
High
Alarm
alerts
user
when
reached
Low
Limit
Low
Alarm
alerts
user
when
reached
Delete
Limits
Clear
the
alarm
limit
values
DELETE
Current
Data
Delete
current
data
All
Data
Delete
all
stored
data
Group
Data
Delete
data
stored
data
plus
alarm
and
calibration
data
MEASUREMENT
VIEW
View
stored
data
in
all
groups
CALIBRATION
Enable
Allow
calibration
access
Disable
Lock
out
the
calibration
mode
Delete
Zero
N
Clear
the
zero
cal.
data
(non
‐
ferrous)
Delete
Zero
F
Clear
the
zero
cal.
data
(ferrous)
NOTE:
Disable
the
Auto
Power
OFF
feature
before
lengthy
programming
to
avoid
inconvenient
automatic
power
down
while
programming.