Setup and Output a process calibration signal
1.
Set the side switch on the left side to the
M/S
position
2.
Turn the function switch to select combined function for the output function required
(
, mA, or V) and related input function will be used simultaneously.
3. Press
the
“SHIFT”
button to shift the buttons for output setting.
4. Use
the
arrow
buttons (
3
/
4
for digit selection,
5
/
6
for value) to set the output
value.
5. Connect
the
OUTPUT
terminals to the device to be tested.
6. Press
the
OUTPUT
button. The process signal will become active.
7. Press
the
OUTPUT
button to return the output to the standby mode.
Figure-1 Quick Output Operation
WARNING
Read "SAFETY INFORMATION" before using this
1. Set switch
to M/S (side)
2. Turn the
function switch to
select combined
function
3. Press SHIFT
for output setting
4. Use the
arrow buttons
to set the
output value.
5. Output
Terminals
6/7.
Press OUTPUT
button to toggle
signal ON/OFF