Description and Button Functions

 

1. LCD Description

 

:  Normal record mode is selected 

:

   Capture record mode is selected. Records the sine 

wave when the measure value exceeds the Capture 

Detection set level.  

Auto power off setting, the meter will automatically 

shut off after 5mins without any button presses.  

:  Average peak measure and record mode selected 

(about 195ms) 

:  

Recording status. After the record mode is set up by the software, press the 

START/STOP button for more than 4 seconds to begin recording. This icon will be 

displayed and indicates the meter is recording data. Press the same button again for 

more than 4 seconds to stop recording.

 

:    Record until full is selected. When the memory is full, the datalogger will stop recording. 

Continuous recording is selected. When the memory is full, the data will overwrite the 

oldest data. 

:   Memory full and recording has stopped. 

:  

Battery sign, when display completely the battery power is full. When there is only a 

frame icon displayed, the battery is exhausted and should be replaced with a new one. 

 

2.  AC adaptor Jack 
3.  ON/OFF Button 
4. MAX/MIN: 

Press the button one time to display the MAX 

(maximum) reading. Press again to display the MIN (minimum) 

reading. A third press will exit MAX/MIN mode. The meter will 

automatically exit MAX/MIN mode in 10 seconds if no button is 

pressed. 

 

5. CH Input Jack:

 Insert the current or voltage sensor to this COM to 

measure and record the data. 

6. PEAK HOLD: 

In the non-PEAK measurement mode, press the 

PEAK HOLD button to display the average Peak value of the 

present value. Press the button again to exit PEAK display. The meter will automatically exit in 

10s if no button is pressed. 

7. START/STOP: 

Press the START/STOP button for one time to pause data. Press again to 

resume. After setting the record mode, press this button for more than 4 seconds and then 

release it after REC is shown on the LCD. The meter will measure and record data according to 

the software setup. 

8. USB Port

 

9. AC Current Clamp Jaw: 

Clamp the jaws around a single 

conductor for AC current measurements to 200A 

10. Jaw Opening Trigger 
11. CH Plug 

Summary of Contents for DL150

Page 1: ...User Manual True RMS AC Voltage Current Datalogger Model DL150 Additional User Manual Translations available at www extech com ...

Page 2: ... the instrument when your hand is wet Do not input over range during measurements Never open the battery cover during measurements Stop using the instrument when there is any structural defect or exposed metal parts Do not install substitute parts or make modifications on the meter Never replace the battery in moist areas Ensure the meter is disconnected and switch off before opening the battery c...

Page 3: ...ull When there is only a frame icon displayed the battery is exhausted and should be replaced with a new one 2 AC adaptor Jack 3 ON OFF Button 4 MAX MIN Press the button one time to display the MAX maximum reading Press again to display the MIN minimum reading A third press will exit MAX MIN mode The meter will automatically exit MAX MIN mode in 10 seconds if no button is pressed 5 CH Input Jack I...

Page 4: ... REC icon will appear in the display indication that recording has begun 6 Press and Hold the START STOP button for four seconds to stop recording The REC icon will disappear from the display NOTE if you STOP the recording you MUST reprogram the meter with the software before recording again 7 Remove the probe from the test circuit connect the datalogger to a PC and use the supplied software to do...

Page 5: ...be used for 60 days or more remove the battery and store separately Note If the clock does not set to the correct time replace the 3V button battery PC Interface and Software Check the software download page of the website www extech com software downloads for the latest version of the PC software Instructions for use and features are detailed in the supplied software HELP utility ...

Page 6: ...D Low battery indication Empty battery symbol appears on the LCD Power supply 4 AAA Batteries Battery life 5 days approximately AC Adaptor 9V 0 5A Standards CAT III 600V Operating Temperature 32 to 122 F 0 to 50 C Operating Humidity 70 RH Current Clamp Jaw opening 0 5 12 7mm Dimensions 4 5 x 2 5 x 1 3 114 x 63 x 34mm Weight 8 7oz 248g Range Accuracy Voltage 40Hz to 1kHz 10VAC to 600 0VAC 2 0 1V Cu...

Page 7: ...ibration for most of our products Contact us for information on calibration and repair availability refer to the contact information below Annual calibrations should be performed to verify meter performance and accuracy Product specifications are subject to change without notice Please visit our website for the most up to date product information www extech com Contact Customer Support Customer Su...

Reviews: