Specifications
Function Range Resolution
Accuracy
DC Voltage
400mV 0.1mV
(0.5% r 2 digits)
4V 0.001V
(1.2% r 2 digits)
40V 0.01V
400V 0.1V
1000V 1V
(1.5% r 2 digits)
AC Voltage
400mV
0.1mV
(2.0% r 10 dgts)
4V 0.001V
(2.0% r 5 dgts)
40V 0.01V
400V 0.1V
1000V 1V
(2.5% r 5 dgts)
DC Current
400
A 0.1
A
(1.0% r 3 digits)
4000
A 1
A
(1.5% r 3 digits)
40mA 0.01mA
400mA 0.1mA
10A 0.01A
(2.5% r 10 digits)
AC Current
400
A 0.1
A
(2.5% r 10 digits)
4000
A 1
A
(1.8% r 5 digits)
40mA 0.01mA
400mA 0.1mA
10A 0.01A
(3.0% r 7 digits)
Resistance
400
0.1
(1.2% r 4 digits)
4k
0.001k
(1.0% r 2 digits)
40k
0.01k
(1.2% r 2 digits)
400k
0.1k
4M
0.001M
40M
0.01M
(2.0% r 3 digits)
Capacitance 40nF
0.01nF
(5.0% r 7 digits)
400nF 0.1nF
(3.0% r 5 digits)
4
F 0.001
F
40
F 0.01
F
100
F 0.1
F
(5% r 5 digits)
Frequency
Hz% Function
5.999Hz 0.001Hz
(1.5% r 1 digits)
59.99Hz 0.01Hz
599.9Hz 0.1Hz
(1.2% r 3 digits)
5.999kHz 0.001kHz
59.99kHz 0.01kHz
599.9kHz 0.1kHz
5.999MHz 0.001MHz
(1.5% r 4 digits)
9.999MHz 0.001MHz
Sensitivity: 0.5V rms <500kHz; 3V rms >500kHz
Duty Cycle
0.1 to 99.9%
Pulse width: 100µs to 100ms, Frequency: 5Hz to 150kHz
NOTE:
Accuracy is stated at 18
o
C to 28
o
C (65
o
F to 83
o
F) and less than 75% RH.
Note:
Accuracy specifications
consist of two elements:
(% reading) – This is the accuracy of the measurement circuit.