MN15A V1.1 02/09
10
General Specifications
Diode Test
Bias voltage: 2.8VDC typical
Continuity Check
Audible signal will sound if the resistance is less than 100
Ω
Input Impedance
1M
Ω
(VDC & VAC)
AC Bandwidth
50 / 60Hz
Display
2000 count (0 to 1999) LCD
Overrange indication
“1___” is displayed
Polarity
Automatic (no indication for positive); Minus (-) sign for
negative
Measurement Rate
2 times per second, nominal
Low Battery Indication
“
”
is displayed
Battery
One (1) 9V battery (NEDA 1604)
Fuses
mA range; 200mA/250V ceramic fast blow
A range; 10A/250V ceramic fast blow
Operating Temperature
32ºF to 122ºF (0ºC to 50ºC)
Storage Temperature
-4
o
F to 140
o
F (-20
o
C to 60
o
C)
Operating Humidity
Max 70% up to 87ºF (31ºC) decreasing linearly to 50% at
122ºF (50ºC)
Storage Humidity
< 80% RH
Operating Altitude
7000 ft. (2000 meters) maximum.
Weight
9.17 oz (255g)
Size
5.2” x 2.6” x 1.5” (132 x 66 x 38mm)
Approvals
CE, ETL
Safety
This meter is intended for indoor use and protected, against
the users, by double insulation per EN61010-1 and IEC61010-
1 2nd Edition (2001) to CAT III 600V; Pollution Degree 2. The
meter also meets UL 61010-1, Second Edition (2004),
CAN/CSA C22.2 No. 61010-1, Second Edition (2004), and UL
61010B-2-031, First Edition (2003)
PER IEC1010 OVERVOLTAGE INSTALLATION CATEGORY
OVERVOLTAGE CATEGORY I
Equipment of OVERVOLTAGE CATEGORY I is equipment for connection to circuits in
which measures are taken to limit the transient overvoltages to an appropriate low level.
Note – Examples include protected electronic circuits.
OVERVOLTAGE CATEGORY II
Equipment of OVERVOLTAGE CATEGORY II is energy-consuming equipment to be
supplied from the fixed installation.
Note – Examples include household, office, and laboratory appliances.
OVERVOLTAGE CATEGORY III
Equipment of OVERVOLTAGE CATEGORY III is equipment in fixed installations.
Note – Examples include switches in the fixed installation and some equipment for
industrial use with permanent connection to the fixed installation.
OVERVOLTAGE CATEGORY IV
Equipment of OVERVOLTAGE CATEGORY IV is for use at the origin of the installation.
Note – Examples include electricity meters and primary over-current protection equipment