MN26T    V4.0   05/05 

Introduction 

Congratulations on your purchase of Extech’s MN26 Autoranging Multimeter. This meter 
measures AC/DC Voltage, AC/DC Current, Resistance, Capacitance, Frequency, Duty Cycle, 
Temperature, Diode Test and Continuity. Proper use and care of this meter will provide many 
years of reliable service.  

Safety 

International Safety Symbols 

This symbol, adjacent to another symbol or terminal, indicates the user must 
refer to the manual for further information. 
 
This symbol, adjacent to a terminal, indicates that, under normal use, 
hazardous voltages may be present 
 
Double insulation

 

 
 

Safety Precautions 

1. 

Improper use of this meter can cause damage, shock, injury or death. Read and 
understand this user manual before operating the meter. 

2. 

Make sure any covers or battery doors are properly closed and secured. 

3. 

Always remove the test leads before replacing the battery or fuses. 

4. 

Inspect the condition of the test leads and the meter itself for any damage before 
operating the meter. Repair any damage before use. 

5. 

Do not exceed the maximum rated input limits. 

6. 

Use great care when making measurements if the voltages are greater than 25VAC 
rms or 35VDC. These voltages are considered a shock hazard. 

7. 

Always discharge capacitors and remove power from the device under test before 
performing Capacitance, Diode, Resistance or Continuity tests. 

8. 

Remove the battery from the meter if the meter is to be stored for long periods. 

9. 

To avoid electric shock, do not measure AC current on any circuit whose voltage 
exceeds 250V AC. 

10.  Voltage checks on electrical outlets can be difficult and misleading because of the 

uncertainty of connection to the electrical contacts. Other means should be used to 
ensure that the terminals are not "live". 

Specifications 

Function Range 

Resolution 

 

Accuracy 

400mV 0.1mV 

±

(0.5% r 2 digits) 

4V 0.001V 
40V 0.01V 
400V 0.1V 

±

(1.0% r 2 digits) 

DC Voltage 
(V DC) 

600V 1V 

±

(1.5% r 2 digits) 

400mV 0.1mV 

±

(2.0% r 30 digits) 

4V 0.001V 
40V 0.01V 
400V 0.1V 

±

(1.5% r 3 digits) 

AC Voltage 
(V AC) 
(40 - 400Hz) 
(40 - 200Hz for 
400mV, 4V ranges) 

600V 1V 

±

(2.0% r 4 digits) 

400

µ

A 0.1

µ

4000

µ

A 1

µ

40mA 0.01mA 
400mA 0.1mA 

±

(1.5% r 3 digits) 

DC Current 
(A DC) 

10A 0.01A 

±

(2.5% r 5 digits) 

Summary of Contents for MN26

Page 1: ...User s Manual MiniTecTM Series Model MN26 Model MN26T includes temperature probe Mini Autoranging MultiMeter...

Page 2: ...damage before use 5 Do not exceed the maximum rated input limits 6 Use great care when making measurements if the voltages are greater than 25VAC rms or 35VDC These voltages are considered a shock haz...

Page 3: ...reading 5 digits 999 9Hz 0 1Hz 9 999kHz 0 001kHz 99 99kHz 0 01kHz 999 9kHz 0 1kHz 1 2 reading 2 digits 9 999MHz 0 001MHz 1 5 reading 4 digits Frequency Sensitivity 0 8V rms min 20 to 80 duty cycle and...

Page 4: ...AA batteries Fuses mA A ranges 0 5A 250V fast blow 10A range 10A 250V fast blow Operating Temperature 32 o F to 122 o F 0 o C to 50 o C Storage Temperature 4 o F to 140 o F 20 o C to 60 o C Relative H...

Page 5: ...and read the voltage on the display AC or DC Current Measurements CAUTION Do not make current measurements on the 10A scale for longer than 30 seconds followed by a 15 minute cool down period Exceedi...

Page 6: ...robe into the sub miniature temperature socket see separate instruction sheet at the end of this manual 2 Turn the rotary switch to the o C or o F position 3 Read the temperature on the display Note A...

Page 7: ...d reference value A reference voltage current etc can be stored and measurements made in comparison to that value The displayed value is the difference between the reference value and the measured val...

Page 8: ...removed to prevent damage to the unit Battery Replacement WARNING To avoid electric shock disconnect the test leads from any source of voltage before removing the battery fuse cover 1 Disconnect the t...

Page 9: ...returned to Extech The sender is responsible for shipping charges freight insurance and proper packaging to prevent damage in transit This warranty does not apply to defects resulting from action of...

Page 10: ...e thermocouple probe Sensor Type Type K thermocouple Chromel Alumnel Temperature range 40 to 482 o F 40 to 250 o C Lead length 39 1m Connector subminiature Note This probe is not UL listed more infor...