Calibration Zero for Pinless Mode 

1. 

Switch the meter to the Pinless mode of operation (REL mode). 

2. 

Ensure that the meter is not near any objects or surfaces. Hold the meter near the 
bottom to avoid contact with the pinless sensor. 

3. 

Press and Hold the MODE button for 3 seconds until rEL appears in the display. 
Release the MODE button and the display will blink for several seconds and then 
switch back to the normal display mode. 

4. 

The display should now read zero indicating that the calibration is complete. 

5. 

If an E32 error is displayed, try the zero calibration again with the pins further away 
from any objects or surfaces. 

 

Calibration Check for Pin Mode 

1. 

Switch the meter to the Pin mode (%WME). 

2. 

The two calibration check points are located in the holes at the top of the protective 
cap. 

3. 

Insert the electrode pins into the two holes at the top of the protective cap to make 
contact with the test circuit. 

4. 

The display should read between 17.5 to 18.5 %WME, indicating that the 
calibration is verified.  

Summary of Contents for MO265

Page 1: ...User Guide Dual Moisture Meter Pro Pin Pinless Moisture Meter with remote probe Model MO265...

Page 2: ...de years of reliable service Features WME wood moisture equivalent pin moisture reading Relative REL pinless moisture reading for non invasive measurement Remote probe measurements Digital LCD readout...

Page 3: ...ated beneath the protective cap Calibration source located at the top of protective cap A tripod mount is located on the rear of the remote probe head LCD Display Description 1 Measurement reading 2 B...

Page 4: ...of the meter The cap can be affixed to the side of the meter using the protective cap holders 6 2 Press the key to turn the meter ON OFF 3 Press the MODE key to select the Wood Moisture Equivalent WM...

Page 5: ...both at the same time 1 Connect the Remote Probe to the right side of the meter using the RJ 11 plug at the end of the Remote Probe cable 2 Remove the protective cap from the remote probe to expose t...

Page 6: ...switch back to the normal display mode 4 The display should now read zero indicating that the calibration is complete 5 If an E32 error is displayed try the zero calibration again with the pins furthe...

Page 7: ...ght arrow button to change the SETTING Use the HOLD button to scroll through the OPTIONS The available OPTIONS and SETTINGS OPTIONS SETTINGS SET UP INFORMATION 0 0 Switches all settings to 0 Factory D...

Page 8: ...ary but never use abrasives or solvents Prevent dirt from accumulating at the electrode pins Specifications Display Backlit LCD digital display and tri color LED bargraph scale Resolution 0 1 Accuracy...

Reviews: