Model PH220 Version 3.1 11/2005 

Specifications 

 

Range (Resolution)

 

Accuracy

 

mV (ORP)

 

± 1999 mV DC (1 mV)

 

± 2mV

 

Temperature 

 

32 to 212

o

F (<100

o

;1

 o

 >100

o

0 to 100

o

C  (0.1°<100°;1°>100°)

 

± 1.0

o

F  

± 0.5

o

C

 

0.00 to 14.00pH (0.01pH)

 

± 0.02pH

 

pH

 

Note 1: Accuracy does not include pH buffer 

temperature coefficient 

Note 2: Accuracy specified from 0.02pH to 13.98pH 

Temperature Compensation

 

32 to 212

o

F (0-100

o

C) Automatic ‘ATC’ 

Power

 

9 volt alkaline battery 

Display

 

Dual LCD display for simultaneous pH and temperature
readings 

Over-range or open indication

 

Displays "

HHH

" for over-range or open input conditions

Displays “

LLL

” for under-range 

pH Calibration points

 

4.00, 7.00, 10.00 (1, 2, or 3 point calibration can be 
performed) 

Auto Power OFF 

10 minutes from last key-press (can be disabled) 

Low Battery Indication

 

Fill bars in the battery display icon           represent 
battery power status. The fewer the bars, the weaker 
the battery power. 

Dimensions / Weight

 

4.25 x 3.00 x 1.18" (108 x 75 x 30mm) / 4.8 oz. (135g) 

Accessories supplied with kit

 

Battery, suction cup holder, clip holder, stick electrode 
for model PH220-S, standard electrode with 39” cable 
for model PH220-C. 

Summary of Contents for PH220A

Page 1: ...evice with tactile push buttons is battery operated and ideal for field use The PH220 simultaneously displays pH and temperature The rubber holster which protects the meter has a magnetic backing for...

Page 2: ...simultaneous pH and temperature readings Over range or open indication Displays HHH for over range or open input conditions Displays LLL for under range pH Calibration points 4 00 7 00 10 00 1 2 or 3...

Page 3: ...play icons are used and are discussed in the appropriate sections of this manual Order of Operations 1 Calibrate the instrument as described later in this manual 2 Determine the test type pH or ORP an...

Page 4: ...nction does not require calibration pH Calibration Procedure 1 Place the electrode into a 7 00 pH buffer solution Press and hold the CAL button until CAL appears in the display 2 The instrument automa...

Page 5: ...g the on off button Wait for the self test SLF tSt to finish 4 Use the MODE button to select mV mode The display will reflect the unit of measure 5 Place the electrode into a sample solution When the...

Page 6: ...time the recall mode is accessed 3 While in the RECALL mode use the STORE RECALL button to scroll through the stored readings The most recent readings will show first LAST IN FIRST OUT 4 Press and ho...

Page 7: ...it over the top of the meter ensure that the electrode is disconnected 3 Pry the battery compartment up from the top of the meter using a screwdriver or coin at one of the two indentations see diagra...

Page 8: ...ine measurement Soak in 0 1 M HCL overnight Deteriorated Gel layer Replace electrode Protein coating on electrode surface Soak in 1 gm Pepsin dissolved in 100ml of 0 1 m HCL for 30 minutes or as neede...

Page 9: ...o sensors and cables If it should become necessary to return the instrument for service during or beyond the warranty period contact the Customer Service Department at 781 890 7440 ext 210 for authori...

Reviews: