Model PH220 Version 3.1 11/2005 

Battery Replacement 

The 9 volt alkaline battery that powers the instrument requires replacing when all battery 
icon ‘fill bars’           are depleted. To replace the battery follow the steps below: 

1. 

Disconnect the electrode from the instrument. 

2. 

Remove the instrument’s protective jacket by pulling it over the top of the meter 
(ensure that the electrode is disconnected). 

3. 

Pry the battery compartment up from the top of the meter using a screwdriver or 
coin at one of the two indentations (see diagram). 

4. 

Replace the 9V battery. 

5. 

Insert the battery compartment into the instrument until the compartment is flush 
with the top of the instrument. The compartment can only be fully inserted in one 
direction and the battery compartment is labeled as to the polarity orientation of the 
battery. 

6. 

Replace the protective holster. 

 

Summary of Contents for PH220A

Page 1: ...evice with tactile push buttons is battery operated and ideal for field use The PH220 simultaneously displays pH and temperature The rubber holster which protects the meter has a magnetic backing for...

Page 2: ...simultaneous pH and temperature readings Over range or open indication Displays HHH for over range or open input conditions Displays LLL for under range pH Calibration points 4 00 7 00 10 00 1 2 or 3...

Page 3: ...play icons are used and are discussed in the appropriate sections of this manual Order of Operations 1 Calibrate the instrument as described later in this manual 2 Determine the test type pH or ORP an...

Page 4: ...nction does not require calibration pH Calibration Procedure 1 Place the electrode into a 7 00 pH buffer solution Press and hold the CAL button until CAL appears in the display 2 The instrument automa...

Page 5: ...g the on off button Wait for the self test SLF tSt to finish 4 Use the MODE button to select mV mode The display will reflect the unit of measure 5 Place the electrode into a sample solution When the...

Page 6: ...time the recall mode is accessed 3 While in the RECALL mode use the STORE RECALL button to scroll through the stored readings The most recent readings will show first LAST IN FIRST OUT 4 Press and ho...

Page 7: ...it over the top of the meter ensure that the electrode is disconnected 3 Pry the battery compartment up from the top of the meter using a screwdriver or coin at one of the two indentations see diagra...

Page 8: ...ine measurement Soak in 0 1 M HCL overnight Deteriorated Gel layer Replace electrode Protein coating on electrode surface Soak in 1 gm Pepsin dissolved in 100ml of 0 1 m HCL for 30 minutes or as neede...

Page 9: ...o sensors and cables If it should become necessary to return the instrument for service during or beyond the warranty period contact the Customer Service Department at 781 890 7440 ext 210 for authori...

Reviews: