Air Flow (Volume CFM/CMM) 

1. 

First measure and note the area of the duct or other passageway under test 

in square feet or 

square meters

. For rectangular ducts, area is calculated using 

Length * Width

; for circular 

ducts use 

3.14 * R

2

.   

2. 

Enter the area value in the SETUP mode (refer to SETUP mode section for details). 

3. 

Select the Air Volume mode on the meter as described earlier using the FUNCTION key. 

4. 

Connect the hot wire probe to the meter’s probe jack (top right of meter) and slide open the 
probe’s protective cover to expose the hot wire sensor. Close the protective cover when the 
probe is not in use. 

5. 

Hold the probe by its handle and allow the air flow to pass through the sensor area. Note that 
the air flow must face the probe’s colored dot. Refer to the meter and probe diagram section. 

6. 

The meter will display the air volume measurement in the upper display area and the selected 
unit of measure (US or EU) in the lower display area (

F-US

 for CFM; 

F-EU

 for CMM). CFM 

represents cubic feet per minute and CMM represents cubic meters per minute. 

Data Hold  

To freeze a measurement on the display, press the HOLD key momentarily. The meter will emit a 
beep, the reading will hold, and the HOLD display icon will switch on. Press the HOLD key again to 
release the display and exit the Data Hold mode returning the meter to the normal operating mode. 

MAX-MIN Readings 

For a given measurement session, this meter can record the highest (MAX) and the lowest (MIN) 
readings for later recall. 
1. 

Press the MAX-MIN key momentarily to access this mode of operation (REC icon appears) 

2. 

The meter is now recording the MAX and MIN readings. 

3. 

Press the MAX-MIN key again to view the current MAX reading (MAX icon appears). The 
reading on the display is now the highest reading encountered since the REC icon was 
switched on (when the MAX-MIN key was first pressed). 

4. 

Press the MAX-MIN key again to view the current MIN reading (MIN icon appears). The 
reading on the display is now the lowest reading encountered since the REC icon was 
switched on (when the MAX-MIN key was first pressed). 

5. 

To exit the MAX-MIN mode, press and hold the MAX-MIN key for at least 1.5 seconds. The 
meter will beep, the REC-MAX-MIN icons will switch off, the memory will clear, and the meter 
will return to the normal operating mode.  

Display Backlight 

To turn the display backlight ON or OFF, press and hold the backlight 

 key for at least 1.5 

seconds. The meter will beep when switching the backlight ON or OFF unless the beeper is 
disabled as described in the Setup Mode section of this user guide. 

Summary of Contents for SDL350

Page 1: ...Hot Wire 5 USER MANUAL Hot Wire Thermo Anemometer with Datalogger Model SDL350...

Page 2: ...nt to another symbol or terminal indicates the user must refer to the manual for further information Description METER AND HOT WIRE PROBE 1 Thermocouple input 2 Hot Wire input plug 3 Display for air v...

Page 3: ...isplay the air velocity measurement upper display area and the air temperature lower display area Changing the Air Velocity Unit of Measure This meter offers five 5 units of measure selections for air...

Page 4: ...eading will hold and the HOLD display icon will switch on Press the HOLD key again to release the display and exit the Data Hold mode returning the meter to the normal operating mode MAX MIN Readings...

Page 5: ...he user SD Card Information Insert an SD card from 1G size up to 16G into the SD card slot at the bottom of the meter With regard to orientation the card should be inserted with the front of the card...

Page 6: ...der is created on the card and named AHB01 Up to 99 spreadsheet documents each with 30 000 readings can be stored in this folder 7 When datalogging begins a new spreadsheet document named AHB01001 xls...

Page 7: ...e Setup mode if no key is pressed in 7 seconds while in the Setup mode 5 The available Setup parameters are listed below Additional detailed information is provided below this list dAtE Set the clock...

Page 8: ...setting 4 Press and hold the SET key for at least 1 5 seconds to exit to the normal operation mode or simply wait 7 seconds for the meter to automatically switch to the normal operating mode Numerical...

Page 9: ...ducts 3 14 R2 is used for circular ducts Note that the value must be set in square feet or square meters therefore if the measurement is made in square inches the area value must be converted to squa...

Page 10: ...e sampling rate to 0 Memory Card SD memory card 1G to 16GB size Temperature Compensation Automatic temperature compensation for the Anemometer function and the type K J thermometer function Data Hold...

Page 11: ...1 F 2 0 C 3 6 F Air Flow CFM CMM Volume Measurement unit Range Resolution CMM m3 minute 0 to 54 000 CMM 0 001 to 1 CMM CFM ft3 minute 0 to 1 907 000 CFM 0 001 to 100 CFM Air Temperature via Type K or...

Reviews: