RS-232/USB PC Interface  

For streaming of data to a PC via the RS232 Output jack, the optional 407001-USB kit (RS232 to 
USB cable and driver CD) along with the 407001 software (available free at 

www.extech.com/sdl350

) are required. 

Setup Mode 

Basic settings at a glance 

To view the current configuration of the meter with regard to time, date, thermocouple type and 
datalogging sampling rate, press the SET/CLOCK ICON key momentarily.  The meter will now 
display the configuration in quick succession. Repeat as necessary to observe all of the information. 

Accessing the Setup mode 

1. 

Press and hold the SET key for at least 1.5 seconds to access the Setup menu. 

2. 

Press the SET key momentarily to step through the available parameters. The parameter type 
is shown on the bottom of the LCD and the current selection for that type is shown above it. 

3. 

When a parameter is displayed that is to be changed, use the arrow keys to change the 
setting. Press the ENTER key to confirm a change. 

4. 

Press and hold the SET key for at least 1.5 seconds to exit the Setup mode. Note that the 
meter automatically switches out of the Setup mode if no key is pressed in 7 seconds while in 
the Setup mode.  

5. 

The available Setup parameters are listed below. Additional detailed information is provided 
below this list:  

dAtE    

Set the clock time (Year/Month/Date; Hour/Minute/Second) 

SP-t   

Set the datalogger sampling rate (1 – 3600 Second) 

PoFF   

Automatic power-off management 

bEEP    

Set beeper sound ON/OFF 

dEC   

Set SD card Decimal character (comma for European format) 

Sd F 

   

SD memory card Format 

t-CF   

Select the Temperature unit of measure to 

C

 or 

F

 

tYPE    

Select the thermocouple type to 

K

 or 

J

 

F-US/F-EU

  

Select 

F-US

 for CFM (cubic feet per minute) or 

F-EU

 for CMM (cubic meters 

per minute) regarding Air Flow (Volume) measurements. 

 ArEA 

 

For CFM/CMM Air Flow (Volume) measurements the user enters the area of 
the duct or other air passageway in units of square feet or square meters. 

Setting the Clock Time  

1. Access 

the 

dAtE

 parameter. 

2. 

Use the arrow keys to change a value. 

3. 

Use the ENTER button to step through the selections. 

4. 

Press and hold the SET button for at least 1.5 seconds to exit to the normal operation mode (or 
simply wait 7 seconds for the meter to automatically switch to the normal operating mode). 

5. 

The clock will keep accurate time even when the meter is switched off. However, if the battery 
expires the clock will have to be reset after fresh batteries are installed. 

Summary of Contents for SDL350

Page 1: ...Hot Wire 5 USER MANUAL Hot Wire Thermo Anemometer with Datalogger Model SDL350...

Page 2: ...nt to another symbol or terminal indicates the user must refer to the manual for further information Description METER AND HOT WIRE PROBE 1 Thermocouple input 2 Hot Wire input plug 3 Display for air v...

Page 3: ...isplay the air velocity measurement upper display area and the air temperature lower display area Changing the Air Velocity Unit of Measure This meter offers five 5 units of measure selections for air...

Page 4: ...eading will hold and the HOLD display icon will switch on Press the HOLD key again to release the display and exit the Data Hold mode returning the meter to the normal operating mode MAX MIN Readings...

Page 5: ...he user SD Card Information Insert an SD card from 1G size up to 16G into the SD card slot at the bottom of the meter With regard to orientation the card should be inserted with the front of the card...

Page 6: ...der is created on the card and named AHB01 Up to 99 spreadsheet documents each with 30 000 readings can be stored in this folder 7 When datalogging begins a new spreadsheet document named AHB01001 xls...

Page 7: ...e Setup mode if no key is pressed in 7 seconds while in the Setup mode 5 The available Setup parameters are listed below Additional detailed information is provided below this list dAtE Set the clock...

Page 8: ...setting 4 Press and hold the SET key for at least 1 5 seconds to exit to the normal operation mode or simply wait 7 seconds for the meter to automatically switch to the normal operating mode Numerical...

Page 9: ...ducts 3 14 R2 is used for circular ducts Note that the value must be set in square feet or square meters therefore if the measurement is made in square inches the area value must be converted to squa...

Page 10: ...e sampling rate to 0 Memory Card SD memory card 1G to 16GB size Temperature Compensation Automatic temperature compensation for the Anemometer function and the type K J thermometer function Data Hold...

Page 11: ...1 F 2 0 C 3 6 F Air Flow CFM CMM Volume Measurement unit Range Resolution CMM m3 minute 0 to 54 000 CMM 0 001 to 1 CMM CFM ft3 minute 0 to 1 907 000 CFM 0 001 to 100 CFM Air Temperature via Type K or...

Reviews: