22
SL355-en-GB_V1.8 8/17
rEPt
The REPT (Repeat) function allows the test to be repeated a pre-programmed
number of times (OFF, 0 to 19).
Press and hold RUN/PAUSE for 3 seconds; the display will flash.
Use ▲ or ▼ to change the setting and then press RUN/PAUSE.
Press ▼ to move to the next parameter.
PAIr
Tests ‘pairs’ are pairs of Start (BEG) and End times. When a test pair is active, a
small ‘L’ appears next to the ‘P’ in ‘Pair’.
To disable the test pair, press RUN/PAUSE and the ‘L’ will switch OFF.
Press ▼ to move to the next parameter.
BEG
Test Start (BEGIN) time (HH:MM:SS);
If the TIMER is ON, tests will begin at the BEGIN time after the RUN/PAUSE button
is pressed.
To change the Start time, press RUN/PAUSE, a pair of digits will flash; P
ress ▲or▼
to change the flashing digits.
Press ◄ or ► to select another pair of digits. When done, press RUN/PAUSE to
store the new Start time.
Press ▼ to move to the next parameter.
END
Test END time (HH:MM:SS)
To change the Test END time, press RUN/PAUSE, a pair of digits will flash; P
ress ▲
or ▼ to change the flashing digits.
Press ◄ or ► to select another pair of digits. When done, press RUN/PAUSE to
store the new End time.
Use the MENU button to change mode of operation.
CLrxx
Access this parameter from the TIMER parameter earlier in this menu.
Clear memory location (Run: 01-
20); Press ▼ to access the parameter below or
press ► to return to the current time display
SUreXX
Are you sure? Press RUN/PAUSE to erase the current memory location.
Use the MENU button to change the mode of operation.