2.1.11
Danger of electricity
The device fulfills protection class 2 (protective isolation without protective earth conductor).
Use only the included power adapter (see the information on the nameplates of the
device and plug the AC adapter). The power supply unit is approved for medical
applications and is matched on the connected load of the device. Using a different
power supply may result in injury to the patient or the user!
In case of obvious damage of the power supply (housing or cable) do not use it! In
case of any damage please contact your distributor.
Do not use the unit in case of excessive moisture in the environment. Otherwise,
there is a risk of electric shock.
To avoid short-circuiting caused by condensing water we recommend acclimatizing
the product after transportation for approximately 3 hours.
2.1.12
Fire danger
Do not use the device in
the vicinity of flammable liquids, gases etc…
2.1.13
Essential performances
The essential performances of the ME equpiment are wavelength, illumination intensity, illumination
time and beam profile. To maintain basic safety and essential performances of the equipment, a
regular calibration is mandatory. See chapter 6.2 for detailed information.