CXL-365 vario
corneal cross-linking
1025_IFU_507536 CXL-365 vario_EN_Rev 09
Page 29 of 49
Figure 32: Please note the plug orientation!
Switch on the UV
–
meter by pressing the POWER button and use the measurement range
20mW/cm².
Figure 33: UV-Meter
Do not use the
„HOLD“
-
und „REC“
-button. This might cause wrong measurement
numbers.
In case you see the battery symbol
on the display of the UV - meter you have to
exchange the battery (see section 6.3). Do not use the UV
–
meter with the battery
sign on the display!
Switch on button
Measurement range
20mW/cm²