CXL-365 vario
corneal cross-linking
1025_IFU_507536 CXL-365 vario_EN_Rev 09
Page 42 of 49
Problem Description
Possible cause
Recommended action
Extraneous light hits the sensor
of the UV - meter
Position measurement adapter
correct on UV-Lampe
Sensor of UV-Meters or UV-
Meter display defect
Please contact distributor for
repair
UV
–
meter out of calibration
(see section 6.2)
Please contact distributor for
repair
Illumination check is out of
tolerances
Loss of power of UV diode
Please contact distributor for
repair
UV-meter out of calibration
Please contact distributor for
recalibration
Out of temperature range
Adjust environmental
temperature to 18
–
28 °C
Aperture isn’t
adjusted to
maximum
Please check the aperture
adjustment.