CXL-365 vario
corneal cross-linking
1025_IFU_507536 CXL-365 vario_EN_Rev 09
Page 13 of 49
2.1.14
Electromagnetic compatibility
The device corresponds to the EMC Class B according to EN 60601-1-2:2015 and
should be operated only in the appropriate environmental conditions.
Portable and mobile RF communications equipment can influence other medical
equipment. Make sure that there is enough distance (especially to other medical
devices, mobile phones etc.) to avoid mutual interference between the devices.
People with active implanted medical devices (e.g. pacemakers) should make sure
that they have sufficient distance to the device while operation.
Use of this equipment adjacent to or stacked with other equipment should be
avoided because it could result in improper operation. If such use is necessary, this
equipment and the other equipment should beobserved to verify that they are
operating normally.
Use of accessories, transducers and cables other than those specified or provided by
the manufacturer of this equipment could result in increased electromagnetic
emissions or decreased electromagnetic immunity of this equipment and result in
improper operation.
Portable RF communications equipment (including peripherals such as antenna cables
and external antennas) should be used no closer than 30 cm (12 inches) to any part of
the ME equipment, including cables specified by the manufacturer. Otherwise,
degradation of the performance of this equipment could result.
Guidance and manufacturer's declaration for electromagnetic compatibility see in chapter 2.1.15.