CXL-365 vario
corneal cross-linking
1025_IFU_507536 CXL-365 vario_EN_Rev 09
Page 23 of 49
Do not operate the device unattended! Monitor the device during the treatment;
monitor the proper functioning of the device and the positioning and spacing of the
device to the patient eye!
Figure 21: fundamental sequence of use
Switching on the device positioned
above the Measuring adapter
The device will make a self-test. After
this the device is showing the message
„SELFTEST READY”
and
„Check Energy”
Make the Illumination Check
Position the device above
the eye of the patient
Push the Start/Stop button
The device is in emission mode now;
the treatment time is running
Fine adjustment of distance and
beam diameter at the beginning
of the treatment
Execute the treatment;
the treatment time depends on
adjusted energy
The device stopped automatically
after expired treatment time and
shows the message
”END OF TREATMENT“
Switching off the device
Push the Start/Stop button to pause
Break
–
Illumination is stopped.
The device will show the message
„STOPPED 3mW”, „STOPPED 9mW”or
„STOPPED 18mW”.
Push the Start/Stop button to continue
Interrupt of the treatment is possible!
Adjust Illumination