4.3
Adjust Illumination
With the rotary switch on the front panel, the Illumination of the device can be set. There are three
selection levels: 3, 9 and 18 mW / cm ². The treatment time is set with the energy.
Figure 26: Adjust energy
4.4
Switching on and Commissioning
Connect the power supply with the included adapter cable. Plug in the adapter cable on the upper
side of the housing and on the other hand plug in the power supply in the wall socket. Make sure
that the cable is not in danger to change the positioning of the device.
Make also sure that the plug is not unplugged accidently. Switch on the device with the on / off
button on the right side of the housing. On the display the software version is shown and a self-test is
initiated, which ends with
„SELFTEST READY”
and
„Check Energy”
on the display. The device is now
ready for use.
Figure 27: On / Off button
On / off button
Adjust Illumination
Power In