9
Routine Maintenance
Before Each Use
•
Visual inspection - make sure that all electrical cords are in good condition and hoses do not leak.
•
Make sure it stands securely on the floor.
•
Check the temperature setting by hand or with a bath thermometer.
Daily Maintenance
•
Clean the tub after every use.
•
Check the cleaning supply at least daily, depending on the frequency of use.
•
Avoid abrasive cleaners which may scratch the gel coat finish of the fiberglass bathtub and the panel.
•
When the tub is not in use, lower it to its lowest position and make sure all valves are closed.
•
Clean the strainers on the patient and cleaning shower handles by rubbing it with your thumb.
•
Visually inspect the tub interior to make sure there are no nicks or cracks which may scratch the patient.
Monthly (On-Site Engineer or Trained Staff)
•
Make sure the connecting hot/cold water hoses are fitted correctly and do not leak.
•
Check all screws and bolts.
•
Check the power connection cord for damages (current & potential equalization).
Annually
•
Maintenance and inspection by QUALIFIED STAFF.
•
Movable components have to be maintained and inspected regularly because they are classified as wear parts
(e.g. motor, fittings, hoses, connections, pipe interrupters).
•
Inspection of electrical installation and equipment according to local requirements should be carried out
annually.
Every 5 Years
•
Exchange/replace all connecting hot/cold water hoses.
After Approximately 12 Years
•
EZ Way recommends that both actuators and control system are replaced.
Maintenance and/or repairs may only be carried out by qualified staff. Improper maintenance and/or repair will
void the warranty.
Please note: Iodine, bromine, and methylene blue can discolor the fiberglass.
The recommended frequency of these maintenance procedures will depend on the hardness of your water supply
and how often the equipment is used. It is good practice to establish these routine procedures on a weekly basis
and then adjust the frequency as your conditions indicate.
All mechanical parts inside the bathtub panel are maintenance free.
The equipment must not be used in
any way that varies from the safety
precautions and operating procedures
detailed in this manual.
WARNING:
The equipment must not be operated
until the user has read and understands
the operating manual.
WARNING: