background image

9

Routine Maintenance

Before Each Use

• 

Visual inspection - make sure that all electrical cords are in good condition and hoses do not leak.

• 

Make sure it stands securely on the floor.

• 

Check the temperature setting by hand or with a bath thermometer.

Daily Maintenance

• 

Clean the tub after every use.

• 

Check the cleaning supply at least daily, depending on the frequency of use.

• 

Avoid abrasive cleaners which may scratch the gel coat finish of the fiberglass bathtub and the panel.

• 

When the tub is not in use, lower it to its lowest position and make sure all valves are closed.

• 

Clean the strainers on the patient and cleaning shower handles by rubbing it with your thumb.

• 

Visually inspect the tub interior to make sure there are no nicks or cracks which may scratch the patient.

Monthly (On-Site Engineer or Trained Staff)

• 

Make sure the connecting hot/cold water hoses are fitted correctly and do not leak.

• 

Check all screws and bolts.

• 

Check the power connection cord for damages (current & potential equalization).

Annually

• 

Maintenance and inspection by QUALIFIED STAFF.

• 

Movable components have to be maintained and inspected regularly because they are classified as wear parts 
(e.g. motor, fittings, hoses, connections, pipe interrupters).

• 

Inspection of electrical installation and equipment according to local requirements should be carried out 
annually.

Every 5 Years

• 

Exchange/replace all connecting hot/cold water hoses.

After Approximately 12 Years

• 

EZ Way recommends that both actuators and control system are replaced.

Maintenance and/or repairs may only be carried out by qualified staff. Improper maintenance and/or repair will 
void the warranty.

Please note:  Iodine, bromine, and methylene blue can discolor the fiberglass.

The recommended frequency of these maintenance procedures will depend on the hardness of your water supply 
and how often the equipment is used. It is good practice to establish these routine procedures on a weekly basis 
and then adjust the frequency as your conditions indicate.

All mechanical parts inside the bathtub panel are maintenance free.

The equipment must not be used in 

any way that varies from the safety 

precautions and operating procedures 

detailed in this manual.

WARNING:

The equipment must not be operated 

until the user has read and understands 

the operating manual.

WARNING:

Summary of Contents for Comfortline II

Page 1: ...01 Rev 7 30 18 EZ Way Comfortline II The Wellness Bath Operator s Instructions EZ Way Inc PO Box 89 Clarinda IA 51632 1 800 627 8940 www ezlifts com WARNING Cancer and Reproductive Harm www P65Warning...

Page 2: ...the back The maintenance hatch at the back can be removed for repair and maintenance work 2 The HiLo bathtub can be adjusted between 650 and 1000 mm 26 and 40 3 The three step adjustable foot support...

Page 3: ...and the thermostat and buttons are all working properly The warranty and liability expire automatically if modifications or repairs are carried out or if components are exchanged or opened without EZ...

Page 4: ...Music controls 7 Handles 8 Foot support slots 9 Drain plug 10 Thermostat with bathtub filling spout and shower 11 Shower handle 12 Hand control UP DOWN 13 Overflow drainage and control to open and clo...

Page 5: ...el and close the thermostat Always check the water temperature by hand or a bath thermometer before the patient gets into the bath Lower the tub to the lowest position with the hand control allowing p...

Page 6: ...may use your usual detergent to clean the bathtub Please make sure you have read the detergent supplier s instructions Daily Use of the Cleaning System 1 Remove the cleaning spray handle 2 from bracke...

Page 7: ...seconds starts at minimum capacity and is then turned up to maximum The LED 1 1 is flashing during waiting time LED Color Light System Option Use The light system touch control is located either at t...

Page 8: ...into the card slot 1 and put the lid to the control box back on The MP3 sound system is ready for use after the bathtub mains has been switched on again Autofill Option The Autofill allows staff to t...

Page 9: ...STAFF Movable components have to be maintained and inspected regularly because they are classified as wear parts e g motor fittings hoses connections pipe interrupters Inspection of electrical install...

Page 10: ...hing each button on the hand control INITIALIZIATION must be carried out if the tub does not raise evenly see separate instruction in the manual Check that the thermostat is working properly and test...

Page 11: ...11 Installation Drawing for HiLo Bathtub Comfortline II Floor drainage Open floor drain or HI pipe DN40 50 pipe coupling Wall drainage HI pipe DN40 50 pipe coupling Figure 9 Installation Diagram...

Page 12: ...PARTICULAR USE OR PURPOSE are expressly excluded EZ Way will not be responsible for any expenses inconvenience or incidental or consequential damages occasioned by the use loss of use or misuse of the...

Reviews: