Battery Installation 

1. Put the two AAA batteries into battery compartment in correct polarities. 
2. Push the battery cover horizontally along the arrow shown as below: 

Notes:   

 Battery polarities should be correctly installed. Otherwise, damage may 

be caused to the device. 

 Please put in or remove batteries in right order, or may cause damage to 

the device bracket. 

 Please remove the batteries if the Oximeter will not be used for a long 

time.   

 

Strap Installation 

1. Thread thinner end of the strap through the loop. 
2. Thread thicker end of the strap through the threaded end before pulling 

it tightly. 

Maintenance and Storage 

1. Replace the batteries in time when low voltage lamp is lighted. 
2. Clean surface of the fingertip oximeter before it is used in diagnosis for 

users. 

3. Remove the batteries inside the battery cassette if the oximeter will not 

be operated for a long time. 

4. It is best to preserve the product in a place where ambient temperatures 

is -10

℃~

50

and relative humidity is 10%-95%. 

5. It is recommended that the product should be kept in a dry environment 

anytime. A wet ambient might affect its lifetime and even might damage 
the product. 

6. Avoid exposure or direct sunlight. 
7. Avoid excessive radioactive infrared rays or ultraviolet rays. 
8. Please follow the law of the local government to deal with used battery. 

Technical Specification 

1. Display Type:   
Color OLED display, 4 display directions 
 
2. SPO2: 
Measurement range: 0%-100% 
Resolution: 1% 
Accuracy: 70%-100%, ±2%; 0%-69% no definition. 
 
3. Pulse Rate: 
Measure range

25BPM -250 BPM 

Resolution: 1bpm, 
Accuracy

  2bpm 

Pulse Intensity: Bargraph Indicator   

 
4. Power Requirements:   
Two AAA alkaline Batteries 
Power consumption: 30mA(Normal)   
Low power indication: 
Battery  Life:  Two  AAA  1.5V,  600mAh  alkaline  batteries  could  be 
continuously operated as long as 24 hours.   
5. Dimension: 
Length: 64mm 
Width: 35mm 
Height: 34mm 
Weight: 57g (including two AAA batteries) 
 
6. Environment Requirements: 
Operation Temperature: 5

40

 

Storage Temperature: -10

50

 

Ambient Humidity: 15%-80%, no condensation in operation. 
10%-93%, no condensation in storage   
 
7. Measurement Performance in Low Perfusion Condition:0.3%.   

Declaration 

EMC of this product complies with IEC60601-1-2 standard. 
The materials which the user can come into contact have no toxicity and no 
action on tissues comply with ISO10993-1, ISO10993-5 and ISO10993-10. 

Guidance  and  manufacturer’s  declaration  –  electromagnetic 
emissions-for all EQUIPMENT and SYSTEMS 

 

Guidance and manufacturer’s declaration – electromagnetic 

emission 

The Pulse Oximeter is intended for use in the electromagnetic 
environment specified below. The customer or the user of the Pulse 
Oximeter should assure that it is used in such an environment. 

Emission 
test 

Complian
ce 

Electromagnetic 

environment 

– 

guidance 

RF 
emissions 
CISPR 11 

Group 1 

The Pulse Oximeter uses RF energy only 
for its internal function. Therefore, its RF 
emissions are very low and are not likely 
to cause any interference in nearby 
electronic equipment. 

RF emission 
CISPR 11 

Class B 

The Pulse Oximeter is suitable for use in 
all establishments, including domestic 
establishments and those directly 
connected to the public low-voltage power 
supply. network that supplies buildings 
used for domestic purposes. 

Possible Problems and resolutions 

Problems 

Possible reason 

Solution 

SPO

2

 or PR 

can not be 
shown 
normally 

1. Finger is not inserted 
correctly 

2. User’s Oxyhemoglobin 

value is too low to be 
measured 

1..Retry by inserting the finger 
2. Try some more times, If you 

can make sure no problem 
is existing in the product. 
Please go to a hospital 
timely for exact diagnosis. 

SPO

2

 or PR 

is shown 
unstably 

1. Finger might not be 

inserted deep enough. 

2. Finger is trembling or 

1. Retry by inserting the finger 
2. Try not to move 

user’s body is in 
movement status. 

The 
Oximeter 
can not be 
powered on 
 

1. Power of batteries might 

be inadequate or not be 
there at all 

2. Batteries might be installed 

incorrectly 

3. The Oximeter might be 
damaged 

1. Please replace batteries 
2. Please reinstall the 
batteries 
3. Please contact with local 
customer service centre 

The screen 
are suddenly 
off 

1. The product is 

automatically powered off   
when no signal is 
detected longer than 8 
seconds 

2. Power quantity of the 

batteries is started being 
inadequate 

1. Normal             
2. Replace the batteries 

 

There  are  no  user-serviceable  parts  inside  the  oximeter.  The 
cover should only be removed by qualified service personnel. 

If  you  are  uncertain  about  the  accuracy  of  any  measurement,  check  the 

user’s  vital  signs  by  alternate  means;  then  make  sure  the  oximeter  is 
functioning correctly. 
 
Do  not  spray,  pour,  or  spill  any  liquid  on  the  oximeter,  its  accessories, 
connectors, switches, or openings in the enclosure as this may damage the 
oximeter. 

Symbol Definitions 

Symbol 

Definition   

 

CE Mark: The Product system conforms to essential 
requirements of the Medical Device Directive 
93/42/EEC.

 

 

Type BF equipment (Refer to IEC 60601-1:1995) 

 

Attention, consult accompanying documents. 

SPO

2

 

Oxygen saturation   

/Min 

Pulse rate 

 

Symbol  for  the  marking  of  electrical  and  electronics 
devices according to Directive 200296/EC. 
The  device,  accessories  and  the  packaging  have  to 
be  disposed  of  waste  correctly  at  the  end  of  the 
usage. Please follow Local Ordinances or Regulations 
for disposal. 
Note: The Oximeter is applied to this regulation. 

Note

The illustration used in this manual may differ slightly from the 

appearance of the actual product. 

 

 

 

 

 