Battery Installation
1. Put the two AAA batteries into battery compartment in correct polarities.
2. Push the battery cover horizontally along the arrow shown as below:
Notes:
Battery polarities should be correctly installed. Otherwise, damage may
be caused to the device.
Please put in or remove batteries in right order, or may cause damage to
the device bracket.
Please remove the batteries if the Oximeter will not be used for a long
time.
Strap Installation
1. Thread thinner end of the strap through the loop.
2. Thread thicker end of the strap through the threaded end before pulling
it tightly.
Maintenance and Storage
1. Replace the batteries in time when low voltage lamp is lighted.
2. Clean surface of the fingertip oximeter before it is used in diagnosis for
users.
3. Remove the batteries inside the battery cassette if the oximeter will not
be operated for a long time.
4. It is best to preserve the product in a place where ambient temperatures
is -10
℃~
50
℃
and relative humidity is 10%-95%.
5. It is recommended that the product should be kept in a dry environment
anytime. A wet ambient might affect its lifetime and even might damage
the product.
6. Avoid exposure or direct sunlight.
7. Avoid excessive radioactive infrared rays or ultraviolet rays.
8. Please follow the law of the local government to deal with used battery.
Technical Specification
1. Display Type:
Color OLED display, 4 display directions
2. SPO2:
Measurement range: 0%-100%
Resolution: 1%
Accuracy: 70%-100%, ±2%; 0%-69% no definition.
3. Pulse Rate:
Measure range
:
25BPM -250 BPM
Resolution: 1bpm,
Accuracy
:
2bpm
Pulse Intensity: Bargraph Indicator
4. Power Requirements:
Two AAA alkaline Batteries
Power consumption: 30mA(Normal)
Low power indication:
Battery Life: Two AAA 1.5V, 600mAh alkaline batteries could be
continuously operated as long as 24 hours.
5. Dimension:
Length: 64mm
Width: 35mm
Height: 34mm
Weight: 57g (including two AAA batteries)
6. Environment Requirements:
Operation Temperature: 5
℃
~
40
℃
Storage Temperature: -10
℃
~
50
℃
Ambient Humidity: 15%-80%, no condensation in operation.
10%-93%, no condensation in storage
7. Measurement Performance in Low Perfusion Condition:0.3%.
Declaration
EMC of this product complies with IEC60601-1-2 standard.
The materials which the user can come into contact have no toxicity and no
action on tissues comply with ISO10993-1, ISO10993-5 and ISO10993-10.
Guidance and manufacturer’s declaration – electromagnetic
emissions-for all EQUIPMENT and SYSTEMS
Guidance and manufacturer’s declaration – electromagnetic
emission
The Pulse Oximeter is intended for use in the electromagnetic
environment specified below. The customer or the user of the Pulse
Oximeter should assure that it is used in such an environment.
Emission
test
Complian
ce
Electromagnetic
environment
–
guidance
RF
emissions
CISPR 11
Group 1
The Pulse Oximeter uses RF energy only
for its internal function. Therefore, its RF
emissions are very low and are not likely
to cause any interference in nearby
electronic equipment.
RF emission
CISPR 11
Class B
The Pulse Oximeter is suitable for use in
all establishments, including domestic
establishments and those directly
connected to the public low-voltage power
supply. network that supplies buildings
used for domestic purposes.
Possible Problems and resolutions
Problems
Possible reason
Solution
SPO
2
or PR
can not be
shown
normally
1. Finger is not inserted
correctly
2. User’s Oxyhemoglobin
value is too low to be
measured
1..Retry by inserting the finger
2. Try some more times, If you
can make sure no problem
is existing in the product.
Please go to a hospital
timely for exact diagnosis.
SPO
2
or PR
is shown
unstably
1. Finger might not be
inserted deep enough.
2. Finger is trembling or
1. Retry by inserting the finger
2. Try not to move
user’s body is in
movement status.
The
Oximeter
can not be
powered on
1. Power of batteries might
be inadequate or not be
there at all
2. Batteries might be installed
incorrectly
3. The Oximeter might be
damaged
1. Please replace batteries
2. Please reinstall the
batteries
3. Please contact with local
customer service centre
The screen
are suddenly
off
1. The product is
automatically powered off
when no signal is
detected longer than 8
seconds
2. Power quantity of the
batteries is started being
inadequate
1. Normal
2. Replace the batteries
There are no user-serviceable parts inside the oximeter. The
cover should only be removed by qualified service personnel.
If you are uncertain about the accuracy of any measurement, check the
user’s vital signs by alternate means; then make sure the oximeter is
functioning correctly.
Do not spray, pour, or spill any liquid on the oximeter, its accessories,
connectors, switches, or openings in the enclosure as this may damage the
oximeter.
Symbol Definitions
Symbol
Definition
CE Mark: The Product system conforms to essential
requirements of the Medical Device Directive
93/42/EEC.
Type BF equipment (Refer to IEC 60601-1:1995)
Attention, consult accompanying documents.
﹪
SPO
2
Oxygen saturation
/Min
Pulse rate
Symbol for the marking of electrical and electronics
devices according to Directive 200296/EC.
The device, accessories and the packaging have to
be disposed of waste correctly at the end of the
usage. Please follow Local Ordinances or Regulations
for disposal.
Note: The Oximeter is applied to this regulation.
Note
:
The illustration used in this manual may differ slightly from the
appearance of the actual product.