Model 422 & Model 422CC Instruction Manual
D01187869
Revision A, December 2015
12
•
An 8BC or larger fire extinguisher to fight electrical and oil fires should be
placed within 50 feet of the consistometer.
Certain components are supported during shipment with wooden blocks, foam
padding, plastic ties, etc. to prevent damage. Open the front doors and remove all
the packing materials that would interfere with the operation of the instrument
before powering the instrument. Some components such as touch screen LCD
monitor and computer control modules are removed from the instrument prior to
shipment and may be in a separate container to prevent damage. These devices
must be reinstalled and reconnected before operating the instrument. Before
operating the instrument, it is a good idea to check the bottom of the consistometer
for loose screws or bolts that may have loosened and/or fallen out during shipment.
This is particularly true for overseas shipments.
Locate the box containing the monitor, reticulating arm and computer and remove
them from their shipping containers. Secure the arm and monitor to the
consistometer by sliding the arm mount onto the rear side of the cabinet mounting
panel as shown in Figure 4-1.
Connect the power cord from the monitor to the A/C outlet on the back of the
instrument. Failure to do so can cause the instrument to malfunction. Alternately,
the touchscreen may be powered using a DC power adapter.
Plugging the touchscreen and printer into a different outlet creates
an undesirable ground loop to occur. This creates an unstable
electrical system.
The control box should be connected to the instrument by connecting the 5-pin
power cord to the back of the instrument frame as well as the supplied USB cable.
The instrument will not operate without this USB cable connected from the
computer to the machine.
It is strongly recommended that a 230V uninterruptable power supply (UPS) be
installed on the instrument. This will condition the power signal and prevent loss of
power to the computer. There are two power connectors on the back of the
electrical box. A UPS may be connected between these power connectors to
maintain consistometer computer operation during short power outages or
interruptions. If a UPS is not used, a jumper cord must be connected between these
two connectors or the unit will not power on. An appropriate jumper cord is
included with the consistometer accessories.
A printer is included with the instrument and may be connected to the computer
control box. Connect the USB connector on the printer cable to any USB input on