Model 422 & Model 422CC Instruction Manual
D01187869
Revision A, December 2015
32
7.2
Stopping a Test
To stop the current test, the user must press the STOP TEST button on the RUN-
TIME MENU screen. Once the test has been stopped, no further data will be
logged and no further temperature or pressure control will be provided. The
software will save the current test in a file on the instrument’s local hard disk. The
default location is C:/Fann/Tests. If auto-shutdown is enabled, the test will be
stopped when the Consistency Alarm value has been reached. This action has the
same effect as the user pressing the STOP TEST button; however it also turns the
motor and heater off and the cooling water on.
Note that there is a 5-10 second delay from the time a
consistency alarm is triggered until automatic shutdown occurs.
This is to prevent noise or spikes in the consistency signal from
triggering automatic shutdown prematurely.