RheoVADR
®
Rheometer Instruction Manual
D00845847
Revision E, September 2014
14
5
Operation
To start a test, add 350 ml of sample to the stainless steel sample cup. The sample
cup has a line that marks 350 ml. A scribed line on the rotor indicates proper
immersion depth.
Damage to the bob shaft bearing may occur if the immersion depth
is exceeded.
If other sample holders are used, the space between the bottom of the rotor and the
bottom of the sample holder should be one-half inch (1.27cm) or greater.
This instrument is operated through a keypad (Figure 5-1). The keypad functions
are summarized in Table 5-1 and described fully in the following sections.
Figure 5-1 RheoVADR
®
Rheometer Keypad