RheoVADR
®
Rheometer Instruction Manual
D00845847
Revision E, September 2014
38
Refer to the preceding table and notes that explain how data is captured and recorded.
Table A-1 API Six Speed & Gel Profile
Step
Speed
(rpm)
Run Time
(s)
Mode
Number of Points
to Record
*
1
600
0
S
1
2
300
0
S
1
3
200
0
S
1
4
100
0
S
1
5
6
0
S
1
6
3
0
S
1
7
600
10
C
2
8
0
10
P
2
9
3
10
P
0.1 sec interval
10
600
10
C
2
11
0
10 min
C
2
12
3
10
P
0.1 sec interval
*
Also shows recording rate. For example, Step 9, data points are recorded every 0.1 sec.
Table A-2 API PV/YP & Gel Profile
Step
Speed
(rpm)
RunTime
(s)
Mode
Number of Points
to Record
*
1
600
0
S
1
2
300
0
S
1
3
600
10
C
2
4
0
10
C
2
5
3
30
P
0.1 sec interval
6
600
10
C
2
7
0
10 min
C
2
8
3
30
P
0.1 sec interval
*
Also shows recording rate. For example, Step 5, data points are recorded every 0.1 sec.