Chapter 6: Maintenance and Troubleshooting
Look in the probe compensation log and compare the X, Y and Z
values. Each value should repeat to within the accuracy of the device.
N
OTE
: If one person is able to repeat better than another person, review
probe compensation procedure. A “green light” does not guarantee a
good probe compensation. It simply indicates that the data collected was
within specification.
Mounting
Most accuracy and repeatability problems can be attributed to improper
mounting of the Quantum. These issues are not always obvious. An apparently
stable mount can actually be moving more than the accuracy of the device,
directly impacting results and causing system performance degradation.
Figure 6-7 Probe Compensation Log
08m80e00_FaroArm_Quantum.book Page 136 Tuesday, August 22, 2017 11:03 AM
Summary of Contents for FaroArm Quantum
Page 2: ...08m80e00_FaroArm_Quantum book Page 2 Tuesday August 22 2017 11 03 AM...
Page 4: ...08m80e00_FaroArm_Quantum book Page 2 Tuesday August 22 2017 11 03 AM...
Page 10: ...08m80e00_FaroArm_Quantum book Page vi Tuesday August 22 2017 11 03 AM...
Page 128: ...08m80e00_FaroArm_Quantum book Page 118 Tuesday August 22 2017 11 03 AM...
Page 154: ...08m80e00_FaroArm_Quantum book Page 144 Tuesday August 22 2017 11 03 AM...
Page 170: ...08m80e00_FaroArm_Quantum book Page 160 Tuesday August 22 2017 11 03 AM...
Page 174: ...164 08m80e00_FaroArm_Quantum book Page 164 Tuesday August 22 2017 11 03 AM...
Page 190: ...08m80e00_FaroArm_Quantum book Page 6 Tuesday August 22 2017 11 03 AM...
Page 200: ...08m80e00_FaroArm_Quantum book Page 6 Tuesday August 22 2017 11 03 AM...
Page 201: ...08m80e00_FaroArm_Quantum book Page 1 Tuesday August 22 2017 11 03 AM...