Page 1: ......
Page 2: ...2 3 3 Remote Cable connection 5 2 3 4 Auto on setting 5 2 4 COMPLETE LIGHT CONNECTION TO THE CEILING 6 2 4 1 Ceiling mounting with Faro applications 6 2 4 2 Installation procedure 7 2 4 3 Electrical d...
Page 3: ...standard IEC 60364 1 and national installation regulations for electrical systems in premises for medical use When installed with ceiling floor or wall applications the device must be installed with a...
Page 4: ...all the complete light Note 1 the fuse must be placed on the phase and not on the neutral The lamp must be installed with a multipolar device to separate it from the supply network meeting the require...
Page 5: ...nsulation 1 85 mm Only use certified terminals and connectors with resistance to flame VW 1 or similar Take care that the device is protected with the fuse as recommended in Tab 1 2 3 3 Remote Cable c...
Page 6: ...n limit switch between the fixed and the swivel arm The switch limits must not be passed over or forced 2 4 1 Ceiling mounting with Faro applications Warnings for electrical danger and suspended masse...
Page 7: ...to the screws 8 and fix them with the nuts 9 and washers 10 G Calculate the proper height of the column 14 according to the Formula L H DHC 1020 mm Cut the exceeding column 14 part on the side were no...
Page 8: ...on the column 14 and fix it with the three screws 19 make sure the bores of the bushing 20 match the screws seat on the column 14 and tighten Simultaneously pull the traction cord to push out the con...
Page 9: ...ments of the power supply Stricly comply with the minimum heights shown below 1 Cap 2 Screw 3 Wall application 4 Wall plugs 5 Electrical connector 6 Connector cover 7 Carter 8 Screws 9 Washer 10 Cable...
Page 10: ...cording to requirements of table 1 The resulting medical system must be declared in accordance with IEC EN 60601 1 by the installer Once the fixing point has been established with reference a the cent...
Page 11: ...ch to screw the nut 4 Screw the 2 safety screws F until the cut A into the brass nut is completely closed 4 take care to leave free space around the nut G and the support S to avoid any friction when...
Page 12: ...strenght 4000 V 2 MOPP between primary and sec Thermal Protection Component built in The medical system must be declared compliant with IEC EN60601 1 by the installation technician or manufacturer No...