En
FARO Laser Scanner Photon – User’s Manual
77 / 87
10
Appendix
10.1
FARO Laser Scanner Photon error messages
•
Scan data buffer overflow
Please check the connection between the operating PC and the
scanner. Make sure there is a functional communication with at least
100Mb/s.
If you get one of the following error messages, please contact you
local customer service team for assistance:
•
Failure message received from motor controllers
One of the horizontal or vertical motors issued an error.
•
Timeout: Motor home position
The horizontal motor couldn’t find its zero index.
•
Internal Error: Invalid parameter
The scanner parameters are inconsistent.
•
Failed to open communication line to scanner
The communication with the sensor module couldn’t be established
•
Distance measurement tolerance exceeded
The white reference area on the base gave inconsistent
measurements. Please check cleanliness of reference area, mirror, and
lenses.
•
Timeout: Line sync signal
The mirror motor couldn’t find its zero index.
•
Motor speed tolerance exceeded
The speed of the mirror motor isn’t uniform
10.2
FARO Laser Scanner Photon data and power cable
The scanner is connected by a combined data and power cable from LEMO
®
.
Plugs -Part-No.
Receptacles - Part-No.
FHG.1B.308.CLAD (90° bend)
EGG.1B.308.CLL
FEG.2B.308.CLAD (straight)
HGG.2B.308.CLL