FARO Photon Laser Scanner – User’s Manual
64/68
Revised: April 10, 2008
© 2008 FARO Scanner Production GmbH
10.3.2.2 Split into several files
On tab
Storage
of FARO Record’s scanner parameter dialog you can select that the
scan data stream is automatically split into several files when a certain amount of
columns has been reached.
Fig.89
Split scan into several files
The minimum number is 1000. Smaller scan files are not supported.
For each file, the scan number will be increased and appended to the scan file base
name. The progress dialog will show the current scan file name and the sequence
number separated with a hash sign. Hash sign and sequence number are not part of the
scan file name.
Please note: The file split option is only available for remote storage.
10.3.2.3 Switch external control mode on
The external control mode tells the scanner to obey the start and stop signals of the
external device. If the scanner isn’t in external control mode, it will start scanning as
soon as it gets ready. In addition, no ready signal and no synchronization signals are
issued.
There is no interface to change the external control mode interactively. The external
control mode can be switched on or off by modifying the scanner parameters manually:
Establish a connection to the scanner with FARO Record, make sure that no scanner
parameter dialog is currently open, and then export the current set of scanner
parameters with the command
Scanner
→
Administration
→
Parameter Exchange
→
Parameter From Scanner
.